Repeatability of Post-operative Refraction in Patients With a Novel CE Marked Monofocal Intraocular Lens

Vienna Institute for Research in Ocular Surgery2 个研究点 分布在 1 个国家目标入组 100 人开始时间: 2020年10月1日最近更新:
适应症

试验速览

阶段
不适用
发起方
入组人数
100
试验地点
2
主要终点
Reproducibility of different refraction methods

研究概览

简要总结

Since 2018, the Tecnis Eyhance ICB00 lens (Johnson&Johnson Vision, AMO Germany GmbH) was introduced, which is a monofocal IOL which has the ability of extended depth of focus improving intermediate vision without the photic phenomena specific for the multifocal IOLs. The Tecnis Eyhance ICB00 is one of our standard IOL, routinely used in our clinic. There is no single outcome measurement that can be thought of as summarizing the efficacy of an IOL, nevertheless visual acuity and refractive outcome are the most important parameters for evaluation. Aim of this study is to determine the comparability and reproducibility of different refraction methods in patients implanted with Eyhance lens and compare the outcomes with ZCB00 lens.

详细描述

This a prospective observational study that would include pseudophakic patients that have undergone cataract surgery 2 to 12 months prior. A written informed consent will be obtained from each patient. Patients with ICB00 or ZCB00 implanted lenses will be included.

Autorefraction measurements will be performed consecutively, each patient will be measured three times with each device. Two independent observers will perform subjective refraction and will be masked to the autorefraction outcome.

We would like to include 50 eyes with ICB00 and 50 eyes with ZCB00 lens implanted. For the estimation of sample size, we used data from Wang et al. (Wang L, Misra M, Palikaris IG, Koch DD. J Cataract Refract Surg 2002; 28: 276-282) that compared the refractive outcome of pseudophakic eyes and determined a mean spherical equivalent of -0.76 + 0.96 D for group I and -0.44 + 1.13 for group II, thus leading to an effect size of 0.2. The effect size was used for ANOVA test for repeated measurements within factor, with alpha < 0.05, power of 80% and a correlation of 0.5 between the two groups gave us an estimated of sample size of 42 patient. To allow patient drop out a number of 50 eyes will be included per lens group (in total 100 eyes).

When different lenses are implanted (one eye received ZCB00 and the other Eyhance), than both eyes of the patient will be included.

研究设计

研究类型
Interventional
分配方式
Non Randomized
干预模型
Parallel
主要目的
Other
盲法
None

入排标准

年龄范围
21 Years 至 100 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • 21 years old or above
  • Operated age-related cataract (pseudophakic patient)
  • Tecnis Eyhance ICB00 or Tecnis ZCB00 IOL implanted within 2 to 12 months prior to study related examination

排除标准

  • Intra- or Postoperative complication
  • patients with severe corneal pathology, severe macular degeneration, severe diabetic retinopathy, etc. that would significantly influence visual outcome
  • nystagmus or pathologies that would affect patient fixation
  • pregnancy- for women in reproductive age a pregnancy test will be performed.

结局指标

主要结局

Reproducibility of different refraction methods

时间窗: 2-12 months

次要结局

  • Comparability between the devices (autorefractometers)(2-12 months)

研究者

发起方
Vienna Institute for Research in Ocular Surgery
申办方类型
Other
责任方
Principal Investigator
主要研究者

Prim. Prof. Dr. Oliver Findl, MBA

Head of the Ophthalmology Department, and Principal Investigator

Vienna Institute for Research in Ocular Surgery