G5® Mobile Continuous Glucose Monitoring System Automated vs. Manual Sensor Applicator

Completed
Conditions
Diabetes Mellitus
Registration Number
NCT02696252
Lead Sponsor
DexCom, Inc.
Brief Summary

Comparison of G5® Mobile Continuous Glucose Monitoring System Accuracy When Using an Automated Applicator vs. a Manual Sensor Applicator

Detailed Description

The objective of the study is to demonstrate the performance, i.e. efficacy of the G5x System with the automated applicator is no worse than that of the commercial Dexcom G4 TM PLATINUM Continuous Glucose Monitoring System (G4 System) with the manual applicator.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
97
Inclusion Criteria
  • Ages 2 or older
  • Diagnosis of Type 1 diabetes or Type 2 diabetes
  • Willing to participate in a clinic session involving venous sampling for evaluation of study end point
Exclusion Criteria
  • Use of acetaminophen
  • Known allergy to medical-grade adhesives
  • Pregnancy
  • Hematocrit outside specification of the study-assigned blood glucose meter

Study & Design

Study Type
OBSERVATIONAL
Study Design
Not specified
Primary Outcome Measures
NameTimeMethod
proportion of blood glucose values, compared between CGM and a lab glucose analyzer that are matched relative to a diabetes accuracy measurement of %20/20 mg/dL7 days
Secondary Outcome Measures
NameTimeMethod

Trial Locations

Locations (4)

Joslin Diabetes Center

🇺🇸

Boston, Massachusetts, United States

AMCR Institute

🇺🇸

Escondido, California, United States

Clinical Trials of Texas

🇺🇸

San Antonio, Texas, United States

Rocky Mountain Diabetes and Osteoporosis Center

🇺🇸

Idaho Falls, Idaho, United States

© Copyright 2025. All Rights Reserved by MedPath