跳至主要内容
临床试验/NCT01469715
NCT01469715已完成2 期

A Clinical Study of the Warm-Up, Accuracy, and Lag Performance of a Glucose Blinding Protein-Based Continuous Glucose Monitor (GBP CGM)

University of Virginia2 个研究点 分布在 1 个国家目标入组 16 人开始时间: 2011年11月最近更新:
适应症

试验速览

阶段
2 期
状态
已完成
入组人数
16
试验地点
2
主要终点
Absolute Relative Difference (ARD)

研究概览

简要总结

The purpose of this study is to assess the accuracy of the Becton Dickenson (BD) Technologies Glucose Binding Protein-Based Continuous Glucose Monitor (GBP CGM) in patients with Type 1 diabetes during low (goal glucose 55 mg/dL), normal (80-140 mg/dL) and high (>180mg/dL) glucose levels over a 24 hour period. This will be achieved by monitoring blood sugar levels when a regular dinner meal is given, when a liquid breakfast meal (BOOST Original containing 41 grams Carbohydrates (CHO), 4 grams fat, 10 grams protein) is given, when subcutaneous insulin is dosed to induce hypoglycemia to a goal of 55 mg/dL, and when a regular lunch meal is given. A Continuous Glucose Monitor (CGM) is an electronic device that measures and displays blood sugar (glucose) levels in the body throughout the day and night. The method being used to detect blood sugar in the investigational Glucose Binding Protein-Based Continuous Glucose Monitor (GBP CGM) is different than the method that is currently in use by commercially available models.

Some sensors cannot tell the difference between glucose (sugar) and other substances such as Tylenol, aspirin or citric acid etc. Because they cannot tell the difference, they may give false readings. The GBP CGM is made to only recognize glucose in the body rather than other substances (e.g., Tylenol, aspirin, citric acid, etc.). As a result, the investigators expect the new GBP CGM to be more accurate at detecting low blood sugar levels than the current devices.

详细描述

Development of accurate continuous glucose monitoring devices is critically important for the maintenance of strict glycemic control without increasing the risk for hypoglycemia and for the successful implementation of an artificial pancreas. Glucose binding protein-based (GBP) continuous glucose monitors offer several potential advantages over commercially available glucose oxidase based sensors. Unlike glucose oxidase based sensors, GBP sensors do not require establishment of an equilibrium reduction/oxidation reaction and are not as susceptible to interference from other compounds because they are engineered to recognize only glucose. As a result, GBP-based glucose sensors are exceptionally accurate at low blood glucose levels and have a faster warm-up time than glucose oxidase based systems. The principal idea of this proposal is to evaluate the accuracy and performance of a second generation glucose binding protein-based continuous glucose monitor (GBP CGM) developed by BD Technologies in subjects with type 1 diabetes, with emphasis on warm-up period after insertion, accuracy in hypoglycemic and hyperglycemic ranges, and lag time over a 24 hour period of sensor use. In order to accomplish this, the investigators will conduct a clinical trial consisting of 25.5-hour studies involving 15 subjects, each wearing one active and one mock (no GBP) device simultaneously during meals and induced hypoglycemia to observe a wide range of glucose values. The investigators envision that the glucose binding protein-based continuous glucose monitor will demonstrate minimal warm-up time after insertion, overall accuracy comparable to commercial glucose oxidase sensors with excellent performance in the hypoglycemic range and a nominal lag time to sensing a change in blood glucose.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Supportive Care
盲法
None

入排标准

年龄范围
21 Years 至 65 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Clinical diagnosis of type 1 diabetes mellitus for ≥1 year. For an individual to be enrolled at least one criterion from each list must be met.
  • Criteria for documented hyperglycemia (at least 1 must be met):
  • Fasting glucose ≥126 mg/dL - confirmed
  • Two-hour oral Glucose Tolerance Tests (OGTT) glucose ≥200 mg/dL - confirmed
  • HbA1c ≥6.5% documented - confirmed
  • Random glucose ≥200 mg/dL with symptoms
  • No data at diagnosis is available but the participant has a convincing history of hyperglycemia consistent with diabetes
  • Criteria for requiring insulin at diagnosis (1 must be met):
  • Participant required insulin at diagnosis and continually thereafter
  • Participant did not start insulin at diagnosis but upon investigator review likely needed insulin (significant hyperglycemia that did not respond to oral agents) and did require insulin eventually and used continually
  • Participant did not start insulin at diagnosis but continued to be hyperglycemic, had positive islet cell antibodies - consistent with latent autoimmune diabetes in adults (LADA) and did require insulin eventually and used continually
  • Use of an insulin pump to treat his or her diabetes for at least six months prior to the study.
  • Actively using a bolus calculator function with the current insulin pump with pre-defined parameters for glucose goal(s), carbohydrate ratio(s), and insulin sensitivity factor(s).
  • Signed informed consent
  • Age ≥21 and <65 years old
  • Body mass index between 19 and 30 kg/m2, inclusive
  • HbA1c ≤11%

排除标准

  • Uncontrolled arterial hypertension (diastolic blood pressure >90 mm Hg and/or systolic blood pressure >160 mm Hg)
  • Impaired hepatic function measured as alanine aminotransferase or aspartate aminotransferase ≥three times the upper reference limit
  • Impaired renal function measured as creatinine >1.2 times above the upper limit of normal
  • Diabetic ketoacidosis in the past 6 months
  • Severe hypoglycemia resulting in a seizure or loss of consciousness in the 12 months prior to enrollment
  • Conditions which may increase the risk of induced hypoglycemia such as known coronary artery disease, congestive heart failure, history of any cardiac disorder or arrhythmia, history of a cerebrovascular event, history of migraines, seizure disorder, syncope, adrenal insufficiency, or neurological disease.
  • Current use of medications containing >4000 mg acetaminophen per day.
  • Current use of L-Monoamine oxidases (MAO) inhibitors.
  • Known microvascular (diabetic) complications (other than diabetic non-proliferative retinopathy),such as history of laser coagulation, proliferative diabetic retinopathy, known diabetic nephropathy (other than microalbuminuria with normal creatinine) or neuropathy requiring treatment
  • Known allergy to eggs
  • Pregnancy, breast-feeding or intention of becoming pregnant
  • Current or recent alcohol or drug abuse by patient history.

结局指标

主要结局

Absolute Relative Difference (ARD)

时间窗: 25.5 hours

ARD=100\*(G_sensor-G_reference)/G_reference Calculated for when patient's G_ref was Normal (70-180 mg/dl), Hyperglycemic (\>180 mg/dl) and Hypoglycemic (\<70 mg/dl) The study data includes 208 paired sensor-YSI plasma glucose readings (G_reference) for each GBP CGM sensor (G_sensor) inserted for 24 hours during hyperglycemic and hypoglycemic challenge conditions. Data pairs will permit the detailed evaluation of sensor performance parameters, including static accuracy metrics such as median and mean absolute deviations and median and mean absolute relative deviation and Point CG-EGA, as well as dynamic parameters, such as warm-up time, trend accuracy (Rate CG-EGA), and sensor lag.

次要结局

未报告次要终点

研究者

申办方类型
Other
责任方
Principal Investigator
主要研究者

Stacey Anderson

Assistant Professor

University of Virginia

研究点 (2)

Loading locations...

相似试验