跳至主要内容
临床试验/NCT03166566
NCT03166566已完成不适用

Laminar Drainage Implant: Safety of a Novel Surgical Treatment for Refractory Glaucoma.

Federal University of Minas Gerais0 个研究点目标入组 15 人开始时间: 2011年3月1日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
15
主要终点
Change for baseline of intra-ocular pressure (Goldmann applanation tonometry)

研究概览

简要总结

Purpose: To assess the feasibility of a surgical technique and present the preliminary safety results of a new glaucoma device devoid of a tube in painful blind eyes.

Methods: Fifteen end-stage glaucomatous eyes without light perception vision were treated with a novel laminar drainage implant. Intraocular pressure was measured preoperatively and up to 24 months after surgery using a Goldmann applanation tonometer. A scale ranging from 0 to 10 was used to evaluate ocular pain. Conjunctival hyperemia, discharge, erosion or retraction, aqueous humor leakage, corneal edema, hyphema, anterior chamber cells and depth, dislocation of the implant, and filtering bleb height were assessed by slit-lamp biomicroscopy. Anterior segment optical coherence tomography was also assessed.

详细描述

Methods: Light perception was tested in the preoperative assessment. The following parameters were evaluated at the pre- and postoperative visits: morning Goldmann applanation tonometry; anterior segment biomicroscopy, gonioscopy, and eye pain intensity. Postoperative visits occurred on days 1, 3, 7, 14, and 30, once a month up to 6 months, and every 6 months from 6 months to 2 years. The criteria for IOP reduction success were 15 mmHg and a 20% reduction compared to the baseline preoperative IOP. Biomicroscopy variables included the presence or absence of hyperemia, discharge, conjunctival erosion and retraction, aqueous humor leakage (Seidel test), corneal edema, hyphema, AC reaction and depth, and device dislocation. A shallow AC was considered a Spaeth classification grade of I or II and a flat AC was considered grade III. A filtering bleb height above the LDI profile was considered elevated; otherwise, it was classified as flat. Gonioscopy was performed preoperatively and postoperatively. Ocular pain intensity was assessed before and after surgery using a verbal Numerical Rating Scale ranging from zero to 10, with zero representing the absence of pain and 10 representing the worst eye pain ever felt before surgery.

Anterior segment optical coherence tomography (AS-OCT) (Visante®, Carl Zeiss Meditec®, Inc., Dublin, CA, USA) was performed postoperatively at the highest resolution (enhanced high-resolution corneal mode).

Reoperations to unblock the internal ostium were not considered failures; thereafter, the day of the second surgery was considered postoperative day 0.

LDI A human LDI was designed by one of the authors (S.J.) after a preclinical trial in rabbits with a similar device. It is a 150-µm-thick foil made of polymethylmethacrylate (PMMA), composed of three functional areas (Fig. 1): 1) the tip, which has a triangular shape and is inserted into the AC angle; 2) the intermediate portion, which has a rectangular shape (4.5 mm long × 4.0 mm wide) and is positioned inside a scleral tunnel; and 3) the posterior plate, which has a round shape (diameter = 12.5 mm) and is positioned between the sclera and conjunctiva. The LDI plate is positioned only 4.5 mm from the corneoscleral limbus and no allograft coverage is necessary. Its hypothetical mechanism of action is similar to those of popular GDIs and shunts the aqueous humor from the AC to the subconjunctival plate. The difference between their designs is the absence of a tube in the LDI. Thus, aqueous humor drainage is supposed to occur in the space between the inner walls of the scleral tunnel and LDI and/or thought is cleft.

LDI Surgery All surgeries were performed under local peribulbar anesthesia by the same surgeon (S.J.).

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Device Feasibility
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • patients aged 18 years or older with a painful blind eye refractory to clinical treatment and an intraocular pressure (IOP) ≥ 30 mmHg.

排除标准

  • eyes with extensive conjunctival scarring, a disorganized anterior segment, or retinal detachment identified by ophthalmoscopy or B-scan ultrasonography.

结局指标

主要结局

Change for baseline of intra-ocular pressure (Goldmann applanation tonometry)

时间窗: on days 1, 3, 7, 14, and 30, once a month up to 6 months, and every 6 months from 6 months to 2 years.

The criteria for IOP reduction success were 15 mmHg and a 20% reduction compared to the baseline preoperative IOP.

次要结局

  • Conjunctival discharge presence/absence change from baseline(on days 1, 3, 7, 14, and 30, once a month up to 6 months, and every 6 months from 6 months to 2 years.)
  • Conjunctival erosion or retraction appearence after surgery(on days 1, 3, 7, 14, and 30, once a month up to 6 months, and every 6 months from 6 months to 2 years.)
  • Corneal edema presence change from baseline(on days 1, 3, 7, 14, and 30, once a month up to 6 months, and every 6 months from 6 months to 2 years.)
  • Ocular pain intensity ratting change from baseline(on days 1, 3, 7, 14, and 30, once a month up to 6 months, and every 6 months from 6 months to 2 years.)
  • Filtering bleg height variation from baseline(on days 1, 3, 7, 14, and 30, once a month up to 6 months, and every 6 months from 6 months to 2 years.)
  • Anterior chamber depth (gonioscopy) change from baseline(on days 1, 3, 7, 14, and 30, once a month up to 6 months, and every 6 months from 6 months to 2 years.)
  • Optic coherence tomography (OCT) assessment qualitative study after surgery(from 6 months to two years after surgery.)
  • Conjunctival hyperemia change comparing to baseline(on days 1, 3, 7, 14, and 30, once a month up to 6 months, and every 6 months from 6 months to 2 years.)

研究者

发起方
Federal University of Minas Gerais
申办方类型
Other
责任方
Principal Investigator
主要研究者

SÉRGIO JACOBOVITZ

co-advisor

Federal University of Minas Gerais