Skip to main content
Clinical Trials/NCT03869840
NCT03869840CompletedNot Applicable

Accuracy of Detection of Methemoglobin With Pulse Oximetry

Nonin Medical, Inc1 site in 1 country13 target enrollmentStarted: March 26, 2019Last updated:
Conditions

Trial Snapshot

Phase
Not Applicable
Status
Completed
Enrollment
13
Locations
1
Primary Endpoint
Validate %MetHb

Study Overview

Brief Summary

The non-invasive measurement of dysfunctional hemoglobins allows physicians to monitor levels in affected individuals without requiring blood samples or laboratory testing. Nonin Medical Inc, has recently developed a multi-wavelength device that measures carboxyhemoglobin and methemoglobin non-invasively. The primary objective of this study is to validate %MetHb and SpO2 accuracy performance of the study device. Evaluations will be for the range of 0 to 15% MetHb with 95-100% SaO2 as assessed by CO-oximetry, and 0-15% MetHb accuracy under conditions of elevated HHb (SaO2 80-100%) as assessed by CO-oximetry, and SpO2 with 80-100% SaO2 during elevated MetHb as assessed by CO-oximetry.

Detailed Description

The goal of this study is to collect data with induced hypoxia, with induced methemoglobinemia, and with induced hypoxia and methemoglobinemia. This will be accomplished in three runs with 24 to 28 plateaus total. The number of plateaus will be dependent on the subject's tolerance of the study procedure. The first run will be strictly induction of hypoxia down to approximately 70% saturation. The second run will be induction of hypoxia down to approximately 80% saturation with mild hypoxia at 7% MetHb. The third run will be induction of hypoxia down to approximately 80% saturation with a target MetHb of 11-15%.

Study Design

Study Type
Observational
Observational Model
Cohort
Time Perspective
Prospective

Eligibility Criteria

Ages
18 Years to 50 Years (Adult)
Sex
All
Accepts Healthy Volunteers
Yes

Inclusion Criteria

  • The subject is male or female.
  • The subject is of any racial or ethnic group.
  • The subject is between 18 years and 50 years of age (self-reported).
  • The subject does not have significant medical problems (self-reported).
  • The subject is willing to provide written informed consent and is willing and able to comply with study procedures.

Exclusion Criteria

  • Has a BMI greater than 31 (calculated from self-reported weight and height).
  • Has had any relevant injury at the sensor location site (self-reported).
  • Has a deformity or abnormalities that may prevent proper application of the device under test (based on visual inspection).
  • Has a known respiratory condition (self-reported).
  • Is currently a smoker (self-reported).
  • Has a known heart or cardiovascular condition (self-reported).
  • Is currently pregnant (self-reported).
  • Is female and actively trying to get pregnant (self-reported).
  • Has a clotting disorder (self-reported).
  • Has Raynaud's Disease (self-reported).
  • Is known to have a hemoglobinopathy such as (anemia, bilirubinemia, sickle-cell anemia, inherited or congenital methemoglobinemia) (self-reported).
  • The subject has a COHb greater than 3% or MetHb greater than 2% (based on the first blood sample analysis).
  • Has taken blood thinners or medication with aspirin within the last 24 hours (self-reported).
  • Has unacceptable collateral circulation from the ulnar artery (based on exam).
  • Has donated more than 300 mL of blood within one month prior to start of study (self-reported).
  • Is unwilling or unable to provide written informed consent to participate in the study.
  • Is unwilling or unable to comply with the study procedures for the primary objectives.
  • Has another health condition which in the opinion of the principal investigator makes him/her unsuitable for testing.

Outcomes

Primary Outcomes

Validate %MetHb

Time Frame: 4 days

Validate %MetHb accuracy over the range of 0-15% MetHb with SaO2 greater than 95% as assessed by CO-oximetry

Secondary Outcomes

  • Validate the %MetHb accuracy(4 days)

Investigators

Sponsor Class
Industry
Responsible Party
Sponsor

Study Sites (1)

Loading locations...

Similar Trials

Accuracy of Detection of Methemoglobin... | Clinical Trial