MedPath

Study Comparing Onyx and TRUFILL in Brain Arteriovenous Malformations (AVMs)

Phase 2
Completed
Conditions
Brain Arteriovenous Malformations
Interventions
Device: Onyx
Device: TRUFILL
Registration Number
NCT00857662
Lead Sponsor
Medtronic Neurovascular Clinical Affairs
Brief Summary

Test whether AVMs treated with Onyx is equivalent to treatment with n-BCA. Success is defined as an AVM size reduction greater than 50%

Detailed Description

Recent advances in the endovascular treatment of arteriovenous malformations (AVMs) have increased the number of patients with brain AVMs for whom embolization therapy may be appropriate. The permanency of obliterated nidi and occurrence of procedural complications are thought to be at least partially influenced by the characteristics of the material used, with liquid agents more likely to reach and occlude the AVM nidus compared to particulate embolic agents.

The only liquid embolic agent approved in the U.S. for the presurgical embolization of AVMs is TRUFILL®. TRUFILL n-butyl cyanoacrylate (n-BCA) is a liquid adhesive that polymerizes into a solid material upon contact with blood fluids or tissue, via an anionic mechanism. TRUFILL Ethiodized Oil is mixed into the n-BCA monomer as a radiopaque polymerizing retardant. TRUFILL Tantalum Powder may also be added for radiopacity. The TRUFILL n-BCA Liquid Embolic System received U.S. FDA premarket approval on September 25, 2000 (P990040) for use in the embolization of cerebral AVMs, when presurgical devascularization is desired.

Onyx™ is a non-adhesive liquid embolic agent comprised of ethylene vinyl alcohol (EVOH) copolymer dissolved in dimethyl sulfoxide (DMSO), and of micronized tantalum powder. Onyx precipitates into a solid on contact with blood fluids, due to rapid diffusion of the DMSO solvent. The Onyx Liquid Embolic System received the European "CE mark" in July 1999, and has been available outside of the U.S. since September 1999 for use in the embolization of AVMs.

The purpose of this randomized-controlled study is to obtain prospective clinical data on the performance of Onyx (investigational device) and TRUFILL (control device) in the presurgical embolization of brain AVMs. Device safety will be assessed by comparing overall and device-related morbidity and mortality. The primary efficacy endpoint is the angiographic reduction in AVM size (volume) achieved. The objective is to demonstrate that Onyx is no worse than TRUFILL within a specified clinical tolerance. Study results will be used to support a premarket approval application for Onyx in the presurgical embolization of brain AVMs.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
117
Inclusion Criteria
  • The patient or patient's guardian understands and will sign the informed consent for the procedure
  • The patient has a confirmed diagnosis of a brain AVM in the cerebral cortex, cerebellum or dura mater as visualized by angiography or cross sectional imaging.
  • The brain AVM has a Spetzler-Martin grade of I, II, III, or IV. If the brain AVM has a Spetzler-Martin grade of I or II, the anticipated benefit of embolization for surgical resection is greater than the risk of the embolization procedure (e.g., patient stability).
  • The patient is a candidate for surgical resection of the AVM post embolization.
  • The patient is clinically and neurologically stable, for a minimum of 24 hours prior to embolization.
  • The patient agrees to have, and is capable of completing, all study-related exams and procedures.
  • Patient of any age.
Exclusion Criteria
  • The patient is pregnant.
  • The patient has a brain AVM with high flow arteriovenous fistulae that the investigator has determined to be unsuitable for embolization.
  • The brain AVM has a Spetzler-Martin grade of V.
  • The patient is participating in another research study involving another investigational device, procedure or drug.
  • The brain AVM has been previously treated with another embolization agent

Study & Design

Study Type
INTERVENTIONAL
Study Design
PARALLEL
Arm && Interventions
GroupInterventionDescription
OnyxOnyx-
TRUFILLTRUFILL-
Primary Outcome Measures
NameTimeMethod
Angiographic reduction in AVM size (volume) of 50% or greater, where angiographic size reduction is defined as the change from the original AVM size prior to any embolization procedure, to the AVM size after the last embolization.Post final embolization
Secondary Outcome Measures
NameTimeMethod
Surgical resection time
Safety will be assessed by the nature and severity of adverse events3 months post discharge
Surgical blood loss

Trial Locations

Locations (1)

UCLA

🇺🇸

Los Angeles, California, United States