Contribution of an Off-the-shelf Carbon Ankle Foot Orthosis (AFO) in the Management of Walking Functional Impairment
- Conditions
- Neuropathy (Disorder)
- Interventions
- Device: Phase A1Device: Phase B1Device: Phase A2Device: Phase B2
- Registration Number
- NCT05131399
- Lead Sponsor
- Thuasne
- Brief Summary
The objective of this study is to assess the superiority in achieving personal goals (set by the patient and the medical team) according to the GAS scale (Goal Attainment Scaling) while wearing the carbon medical device Ankle Foot Orthosis (AFO) compared to a standard plastic ankle foot orthosis.
- Detailed Description
Foot drop is a symptom observed in various pathologies including neuro-muscular pathologies of peripheral or central origin. It is a walking disability, causing a deterioration in the patients autonomy and quality of life.
The current medical strategy focuses on various off-the-shelf or custom made devices, depending on the walking functional impairment and the characteristics of foot deformities.
The objective of this study is to assess the superiority in achieving personal goals (set by the patient and the medical team) according to the GAS scale (Goal Attainment Scaling) while wearing the carbon medical device ankle foot orthosis compared to a standard plastic ankle foot orthosis.
Recruitment & Eligibility
- Status
- UNKNOWN
- Sex
- All
- Target Recruitment
- 20
- Patient with a foot levator muscles impairment (single or bilateral drop foot) and / or instability of the back of the foot, associated with a deficit of the triceps
- Patient with inability to stand on tip-toe on one foot
- Patient having signed a free and informed consent
- Patient affiliated or entitled to a social security scheme
- Patient with fixed varus-equine foot
- Patient with an unstable skin condition (whether or not treated with additional treatment that may change the footwear)
- Patient with a a foot levator muscles impairment with a deficiency of the quadriceps and hamstrings (knee involvement)
- Patient with uncontrolled spasticity
- Patient for whom foot surgery is planned during the study
- Patient with one of the contraindications to the use of carbon medical device, indicated in the instructions for use
- Patient with one of the contraindications to the use of the standard plastic ankle foot orthosis, indicated in the instructions for use
- Patient with orthopedic comorbidities affecting the musculoskeletal system
- Pregnant woman
- Patient with major cognitive impairment incompatible with participation in a clinical trial
- Patient participating in another clinical investigation conducted to establish the compliance of a MD impacting the evaluation criteria
- Patient who cannot be followed for 12 weeks
- Vulnerable patient according to article L1121-6 of the public health code, persons subject to a measure of judicial protection or unable to consent freely
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SEQUENTIAL
- Arm && Interventions
Group Intervention Description Self-comparison Phase B1 Patients will be fitted with 2 types of ankle foot orthoses, alternately, divided into 4 phases, in the following order: Phase A1: Standard Plastic ankle foot orthosis Phase B1: carbon medical device ankle foot orthosis Phase A2: Standard Plastic ankle foot orthosis Phase B2: carbon medical device ankle foot orthosis Each patient is his own comparator. For all patients, the total duration of the study will be the same, and will correspond to 12 weeks. Self-comparison Phase A2 Patients will be fitted with 2 types of ankle foot orthoses, alternately, divided into 4 phases, in the following order: Phase A1: Standard Plastic ankle foot orthosis Phase B1: carbon medical device ankle foot orthosis Phase A2: Standard Plastic ankle foot orthosis Phase B2: carbon medical device ankle foot orthosis Each patient is his own comparator. For all patients, the total duration of the study will be the same, and will correspond to 12 weeks. Self-comparison Phase B2 Patients will be fitted with 2 types of ankle foot orthoses, alternately, divided into 4 phases, in the following order: Phase A1: Standard Plastic ankle foot orthosis Phase B1: carbon medical device ankle foot orthosis Phase A2: Standard Plastic ankle foot orthosis Phase B2: carbon medical device ankle foot orthosis Each patient is his own comparator. For all patients, the total duration of the study will be the same, and will correspond to 12 weeks. Self-comparison Phase A1 Patients will be fitted with 2 types of ankle foot orthoses, alternately, divided into 4 phases, in the following order: Phase A1: Standard Plastic ankle foot orthosis Phase B1: carbon medical device ankle foot orthosis Phase A2: Standard Plastic ankle foot orthosis Phase B2: carbon medical device ankle foot orthosis Each patient is his own comparator. For all patients, the total duration of the study will be the same, and will correspond to 12 weeks.
- Primary Outcome Measures
Name Time Method Comparison of the ability to achieve personal goals while wearing carbon or standard plastic ankle foot orthosis 3 times a week during 12 weeks Personal goals achievement is evaluated via the Goal attainment scaling (GAS). Before starting the study, three personal objectives are defined according to patients needs and investigator review.
Raw scores ranging from -3 (degradation) to +2 (maximum achievement of the objective) are defined:
"-3" corresponding to degradation "-2" corresponding to the initial state "-1" corresponding to a partial objective achievement "0" corresponding to the total objective achievement "+1" corresponding to a better objective achievement "+2" corresponding to the maximum objective achievement. The GAS is filled 3 times a week during 12 weeks.
- Secondary Outcome Measures
Name Time Method Comparison of patients fatigue while wearing carbon or standard plastic ankle foot orthosis Visits V0 (Day 0), V1 (between week 2 and 3), V2 (between week 5 and 6), V3 (between week 8 and 9) and V4 (Week 12) Patient's fatigue perception is assessed by the patient via the Borg Rating of Perceived Exertion (RPE) scale following mobility and walking perimeter tests The effort rating is based on a scale of 6 to 20 (6, which is "no effort", to 20, which is "maximum effort").
Comparison of patients stability while wearing carbon or standard plastic ankle foot orthosis Visits V0 (Day 0), V1 (between week 2 and 3), V2 (between week 5 and 6), V3 (between week 8 and 9) and V4 (Week 12) Patient's stability is assessed by 2 means:
1. via the Timed Up and Go test : time (in seconds) to realize the test. The patient should stand up from a chair, walk a distance of three meters, turn around, then walk back to the chair and sit down. These tasks should be performed at a comfortable and safe speed.
2. via the number of falls recorded during each phase and their contextComparison of patients walking perimeter while wearing carbon or standard plastic ankle foot orthosis Visits V0 (Day 0), V1 (between week 2 and 3), V2 (between week 5 and 6), V3 (between week 8 and 9) and V4 (Week 12) Walking perimeter is evaluated via the Six-minute walk test (6MWT), that measures the distance (in meters) during a 6 minutes walk.
Comparison of patients disability evolution according to the investigator while wearing carbon or standard plastic ankle foot orthosis Visits V0 (Day 0), V1 (between week 2 and 3), V2 (between week 5 and 6), V3 (between week 8 and 9) and V4 (Week 12) Doctors' opinion on the Patients disability evolution is measured by the Clinical Global Impression - Improvement (CGI-I) questionnaire.
The scale has 7 levels of answer from "very strongly improved" which is the better outcome (= 1 point) to "very strongly aggravated" which is the worst outcome (= 7 points) .Intermediate levels are: " Significantly improved "(= 2 points) ; " Slightly improved " (= 3 points) ; " No improvement " (=4 points) ; " Slightly worsen " (=5 points) ; " Seriously worsen" ( = 6 points).
The highest score corresponds to the most impairment in quality of life.Comparison of patients disability evolution according to the patient while wearing carbon or standard plastic ankle foot orthosis Visits V0 (Day 0), V1 (between week 2 and 3), V2 (between week 5 and 6), V3 (between week 8 and 9) and V4 (Week 12) Patient's opinion on his disability evolution is measured by the Global Impression of Change (PCI-C).
The scale has 7 levels from "no change or condition has got worsed"(worse outcome) to "a great deal better, and a considerable improvement that has made all the difference ( better outcome)". Intermediate levels are: "almost the same, hardly any change at all"; "a little better, but no noticeable change", "somewhat better, but the change has not made any real difference"; "moderately better, and a slight but noticeable change"; "better, and a definite improvement that has made a real and worthwhile difference.Comparison of pain, gait and balance evolution according to the patient while wearing carbon or standard plastic ankle foot orthosis Visits V0 (Day 0), V1 (between week 2 and 3), V2 (between week 5 and 6), V3 (between week 8 and 9) and V4 (Week 12) Pain level is measured by Visual Analogic Scale (VAS) completed by the patient : 0 corresponds to no pain (better outcome) and 10 to maximum pain (worst outcome).
Natural gait and confidence in walking are measured by Visual Analogic Scale (VAS) completed by the patient : 0 corresponds to no natural gait or no confidence (worst outcome) and 10 to better natural gait and better confidence (better outcome).
Balance is measured by Visual Analogic Scale (VAS) completed by the patient : 0 corresponds to no balance (worst outcome) and 10 to maximum balance (better outcome).Comparison of tolerance of carbon and standard plastic ankle foot orthosis Visits V0 (Day 0), V1 (between week 2 and 3), V2 (between week 5 and 6), V3 (between week 8 and 9) and V4 (Week 12) Tolerance is measured by reported adverse events. Number and type of serious and non-serious Adverse Device Effects (ADE) are reported.
Patients satisfaction with carbon and standard plastic ankle foot orthosis V4 at 12 weeks Patients satisfaction is evaluated at the end of phase B2 regarding the use of the study devices via the "Technology" part of the evaluation towards a technical aid questionnaire (8 questions) and a specific questionnaire.
Comparison of patients mobility while wearing carbon or standard plastic ankle foot orthosis Visits V0 (Day 0), V1 (between week 2 and 3), V2 (between week 5 and 6), V3 (between week 8 and 9) and V4 (Week 12). Mobility is evaluated via staircase tests: Time (in minutes) to go up and down 10 steps
Comparison of acceptability and compliance with carbon and standard plastic ankle foot orthosis Visits V0 (Day 0), V1 (between week 2 and 3), V2 (between week 5 and 6), V3 (between week 8 and 9) and V4 (Week 12) Cumulative wearing time of the devices under study (indicated by a thermal button inserted into the device) is collected during each phase. Specific questions on compliance are also asked by the investigator to the patient during the various visits.
Trial Locations
- Locations (8)
Ch Les Capucins
🇫🇷Angers, France
Chu Dijon
🇫🇷Dijon, France
CH LAVAL
🇫🇷Laval, France
CHU Nantes
🇫🇷Nantes, France
CHU Lille
🇫🇷Lille, France
centre Louis PIERQUIN
🇫🇷Nancy, France
Hôpital Rothschild
🇫🇷Paris, France
CHU Strasbourg
🇫🇷Strasbourg, France