User Performance of the CONTOUR NEXT and CONTOUR TV3 Blood Glucose Monitoring System (BGMS)
- Conditions
- Diabetes Mellitus
- Interventions
- Device: Contour Next and Contour TV3 BGMS testing by subjects with and without Diabetes
- Registration Number
- NCT03730480
- Lead Sponsor
- Ascensia Diabetes Care
- Brief Summary
This study will collect lay user ("subjects") and healthcare professional information on the performance of two different blood glucose monitors, the CONTOUR NEXT and the CONTOUR TV3, in people with and without diabetes. Each subject will use both meters, at different times and in different order, using a fingerstick(s) to collect meter samples. Meter results will be compared to standardized test system to assess the accuracy of the meters' results. Additionally, a blood sample will be drawn from the subject's arm by a trained healthcare professional for testing on both BGMS. Subjects will also be asked to complete a questionnaire to rate their experience using the meters and to evaluate the meter's Instructions for Use. After all testing and the first questionnaire have been completed, the study staff will demonstrate only to subjects with diabetes several features of the meters that would not necessarily be experienced during the meter testing. These subjects will then answer questions about the features in a second questionnaire.
- Detailed Description
This clinical trial will assess the performance (accuracy) of the CONTOUR NEXT and the CONTOUR TV3 monitoring systems ("BGMS") by lay users enrolled as subjects in the study, and by health care professionals. According to a randomization schedule, subjects will be assigned to use either the CONTOUR NEXT or CONTOUR TV3 meters first and then switch and use the second meter, performing the same testing with both meters. They will learn to use the assigned meter by reading the corresponding User Guide (UG) and Quick Reference Guide (QRG). Each subject will perform one fingerstick self-test a palm (AST) test and the study staff will perform a fingerstick test on the subject. All subject self-test results and study staff results will be compared to results from the reference analyzer from a deeper fingerstick. Hematocrit will be measured for all subjects. After testing, all subjects will then complete a questionnaire (Questionnaire 1) to provide feedback on the meter system and instructions for use (UG and QRG). Subjects with diabetes will also have their glucose tested via venipuncture. After all testing is completed on these subjects, the study staff will demonstrate several features of the meter system that would not necessarily be experienced during the brief accuracy testing. These subjects with diabetes will be given a second questionnaire (Questionnaire 2) to provide feedback about the new features of the system. After completing the study procedures for one BGMS, all subjects will then repeat all of these procedures using the second BGMS, except the hematocrit and venipuncture procedures.
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 372
- Males and females, 18 years of age and older
- Ability to speak, read and understand English. Subjects must demonstrate ability to read a paragraph from the first page of the User Guide to qualify for the study.
- Willing to complete all study procedures.
- Hemophilia or any other bleeding disorder.
- Pregnancy (self-reported).
- Physical, visual or neurological impairments that would make the person unable to perform testing with the BGMS.
- Previous participation in a blood glucose monitoring study using the Ascensia CONTOUR NEXT or CONTOUR TV3 BGMS.
- Working for a medical laboratory, hospital or other clinical setting that involves training on or clinical use of blood glucose monitors.
- Working for a competitive medical device company, or having an immediate family member or someone who is not a family member but is living within the household of someone who works for such a company.
- A condition, which in the opinion of the investigator or designee, would put the person or study conduct at risk.
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SEQUENTIAL
- Arm && Interventions
Group Intervention Description Subjects with and without Diabetes test Contour Next and Contour TV3 Contour Next and Contour TV3 BGMS testing by subjects with and without Diabetes All subjects test CONTOUR NEXT BGMS and CONTOUR TV3 BGMS
- Primary Outcome Measures
Name Time Method Fingerstick Contour Next BGMS and Contour TV3 Results Compared to the Reference Analyzer 1 day Number of Participants with self fingerstick results from the Contour Next BGMS and Contour TV3 BGMS that are within 15% of results from the Reference Analyzer. Results from the two meter systems will not be compared to each other.
- Secondary Outcome Measures
Name Time Method Subjects' Responses to 8 Statements Regarding Experience With Contour Next BGMS and Contour TV3 BGMS 1 day A questionnaire was given to participants with 8 statements about the BGMS instructions for use and ease of use of the meter system. These statements were rated by subjects on a Likert Scale with ratings of 1=strongly disagree, 2=disagree, 3=neutral, 4=agree, 5=strongly agree. Participants reporting ratings greater than or equal to 3 are reported. The statements are:
1. I find it easy to do a fingerstick blood test with this meter.
2. The meter display is easy to see and read.
3. It was easy to understand my test results.
4. I like the overall meter design.
5. I find the meter easy to use.
6. The instructions (User Guide and Quick Reference Guide) were easy to understand.
7. The instructions clearly explain how to run a test.
8. The instructions clearly explain what to do if an error message is displayed by the meter.
Note that results from the two meter systems will not be compared to each other.
Trial Locations
- Locations (2)
AMCR Institute
🇺🇸Escondido, California, United States
Ranier Clinical Research Center, Inc.
🇺🇸Renton, Washington, United States