Rapid Diagnostics for HIV and Hepatitis
- Conditions
- HIV InfectionsHepatitis B InfectionsHepatitis C Infections
- Interventions
- Device: Multiplo HBc/HIV/HCV and Reveal HBsAg
- Registration Number
- NCT02190305
- Lead Sponsor
- MedMira Laboratories Inc.
- Brief Summary
The purpose of this study is to determine the efficacy of two rapid diagnostic tests in plasma, venipuncture whole blood, and fingerstick whole blood. The clinical performance of Multiplo HBc/HIV/HCV will be determined by comparing the results with patient infected status for HIV-1/2 (human immunodeficiency viruses 1 and 2), HBV (hepatitis B virus) and HCV (hepatitis C virus). The clinical performance of Reveal HBsAg will be determined by comparing the results with patient infected status for HBV.
Subject participation in the study will consist of a single one-hour visit, at which time blood samples will be drawn for testing with the investigational devices and with approved comparator assays. The test results, which are the outcome of the study, will be obtained only once, at the time of this visit.
- Detailed Description
The objective of this study is to determine the performance characteristics of two rapid diagnostic tests.
Approximately 3000-4500 subjects in total will be enrolled in the study into four different study populations. In Population 1, which is "at risk" group of about 2000-3000 subjects, roughly 2/3 will be individuals who are at risk of infection with HIV or hepatitis B or C, or who have signs or symptoms of hepatitis. Approximately 500 of these individuals are expected to be known HIV-positive individuals.
The remaining three study populations will be comprised of individuals with known infection with HIV (Population 1A, \~500 individuals), HBV (Population 1B, \~500 individuals), and HCV (Population 1C, \~500 individuals).
Both Multiplo HBc/HIV/HCV and Reveal HBsAg will be used to test finger stick whole blood, venous whole blood and plasma samples from each subject in Population 1. Plasma samples (repository or fresh) will be used to fulfill the requirements for Populations 1A, 1B, and 1C.
For all enrolled subjects, the plasma sample will be tested with the following FDA-approved assays: EIA for anti-HIV-1/2, anti-HCV, anti-HBc total, anti-HBc IgM, anti-HBs, and HBsAg. The efficacy of Multiplo will be determined by comparing the results with patient infected status for HIV-1/2, HBV and HCV. The efficacy of Reveal will be determined by comparing the results with patient infected status for HBV.
The primary analysis will involve comparison of Multiplo HBc/HIV/HCV and Reveal HBsAg results for each of the test markers (anti-HIV-1/2, anti-HCV, anti-HBc, and HBsAg), in each sample matrix (finger stick whole blood, venous whole blood and plasma) with the patient infected status as determined by separate algorithms for HIV, HBV, and HCV. The percent positive and percent negative agreement will be determined relative to patient infected status for each marker and sample type, with corresponding two-sided 95% confidence intervals.
Recruitment & Eligibility
- Status
- WITHDRAWN
- Sex
- All
- Target Recruitment
- Not specified
All subjects:
- 13 years of age or older
- able to sign consent form (or parent/guardian sign consent plus subject sign assent for ages 13-18)
- willing to complete the risk assessment questionnaire, and
- willing to provide blood samples: finger stick sample (2 drops), 1 serum separator tube, 2 anticoagulant (K2 EDTA) tubes.
Population 1 - Diagnostic Trial "At Risk" Group (n = 2000-3000)
Subjects must meet one or more of the following inclusion criteria to be enrolled in Population 1:
- Have at least one risk factor for infection with hepatitis B or C,
- Have at least one risk factor for infection with HIV,
- Present with signs and symptoms of hepatitis,
- Be a known HIV-positive individual (previous positive HIV test result, may comprise up to 500 individuals)
Population 1A - Additional known HIV-positive individuals (n = ~500)
Subjects must:
• be a known HIV-positive individual (previous positive HIV test result).
Population 1B - Additional known HBV-positive individuals (n = ~500)
Subjects must:
• be known to have current or past HBV infection, as evidenced by HBV serology results.
Population 1C - Additional known HCV-positive individuals (n = ~500)
Subjects must:
• be known to be infected with HCV, as evidenced by previous positive HCV test results.
- Subjects do not meet the inclusion criteria
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Arm && Interventions
Group Intervention Description Diagnostic: Multiplo HBc/HIV/HCV + Reveal HBsAg Multiplo HBc/HIV/HCV and Reveal HBsAg Subjects tested with investigational devices and approved comparator assay algorithms for HIV and hepatitis B and C.
- Primary Outcome Measures
Name Time Method Clinical performance of Reveal HBsAg and Multiplo HBc/HIV/HCV At single study visit For each enrolled subject, Multiplo HBc/HIV/HCV and Reveal HBsAg devices will be used to test finger stick whole blood, venipuncture whole blood, and plasma samples. Plasma will also be tested using separate algorithms of FDA approved assays to determine patient infected status for HIV-1/2, hepatitis B, and hepatitis C.
The performance of Reveal HBsAg will be determined relative to patient infected status for hepatitis B for each matrix assessed. The performance of Multiplo HBc/HIV/HCV will be determined relative to patient infected status for each of hepatitis B, HIV-1/2, and hepatitis C.
- Secondary Outcome Measures
Name Time Method
Trial Locations
- Locations (4)
Keck School of Medicine - LAUSC
🇺🇸Los Angeles, California, United States
SCFLD Hepatology Diagnostic Research Laboratory
🇺🇸Miami, Florida, United States
NYC Dept Health & Mental Hygiene
🇺🇸New York, New York, United States
Minneapolis Medical Research Foundation
🇺🇸Minneapolis, Minnesota, United States