跳至主要内容
临床试验/NCT03300037
NCT03300037已完成不适用

HYpopnea and Apnea Detection and Treatment Performance of a New cardiOreSpiratory Holter Monitor

LivaNova5 个研究点 分布在 1 个国家目标入组 46 人开始时间: 2014年9月1日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
LivaNova
入组人数
46
试验地点
5
主要终点
Performance of real-time detector of apnea and hypopnea: sensitivity and positive predictive value of the detector will be calculated when comparing automatically detected respiratory events to gold-standard PSG scoring of apnea/hypopnea events.

研究概览

简要总结

The HYPNOS study is an acute feasibility study to validate a new device for automatic detection and treatment of sleep apnea syndrome. The system is known as the PASITHEA system and it will perform the following tasks:

  • Detect apneas and hypopneas in real time, based on physiological signals acquired with a cardiorespiratory holter (nasal air flow, heart rate and blood oxygen saturation (SaO2))
  • Upon detection of apnea or hypopnea events, trigger kinesthetic stimulation of mechanoreceptors of the skin close to the mastoid bone

The primary objective is to verify that the PASITHEA system is able to detect apneas and hypopneas reliably.

The main secondary objective of the study is to assess the effect of kinesthetic stimulation on reducing the number of sleep respiratory disorders.

Another objective of the study is to verify the safety of the PASITHEA system.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • above 18 years.
  • Patient with diagnosed severe obstructive sleep apnea syndrome (AHI > 30/h and > 80% of obstructive events diagnosed on polysomnography or polygraphy performed less than 6 months ago )
  • Patient free from CPAP treatment or any other SAS treatment for at least 3 months
  • Patient with stable psychotropic medication since the diagnostic polysomnography or polygraphy exam
  • Patient has signed the informed consent

排除标准

  • Patients sleeping less than 4 hours per night
  • Patients suffering from respiratory failure or periodic breathing (cheynes stokes)
  • Obese patient (BMI>40kg/m2)
  • Diabetic patient with autonomic dysfunction
  • Vulnerable patient in accordance with article L1121-6 of CSP
  • Patient with progressive malignant pathology
  • Patient already involved in another clinical study that could affect the result of this study
  • Pregnant patient

结局指标

主要结局

Performance of real-time detector of apnea and hypopnea: sensitivity and positive predictive value of the detector will be calculated when comparing automatically detected respiratory events to gold-standard PSG scoring of apnea/hypopnea events.

时间窗: 1 night

During the night of the recording, the PASITHEA system and the PSG will be positioned in parallel. PSG-scored events will be matched with the output of the real-time detector: a true positive event is defined as an event detected less than 10 seconds before or within the duration of a PSG-annotated events. Apneas detected as hypopneas (and reciprocally) will be regarded as true positives. Sensitivity and positive predictive value will be calculated as per conventional definintion and the 9% confidence interval will be calculated using the Wilson score.

次要结局

  • Safety: number of adverse events with treatment ON or OFF as defined by ISO 14155, rev 2011(1 night)

研究者

发起方
LivaNova
申办方类型
Industry
责任方
Sponsor

研究点 (5)

Loading locations...

相似试验