CAPTure Information Via Automatic Threshold Evaluation
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 216
- 试验地点
- 36
- 主要终点
- Primary Safety Endpoint is to Evaluate the System-related Complication-free Rate
研究概览
简要总结
This study will evaluate the PaceSafe Right Ventricular Autothreshold (RVAT) and Left Ventricular Autothreshold (LVAT) features for AUTOGEN Cardiac Resynchronization Therapy Defibrillator (CRT-D) devices.
详细描述
The objective of the CAPTIVATE Clinical Study is to gather data to establish the safety and effectiveness of the PaceSafe Right Ventricular Autothreshold (RVAT) and Left Ventricular Autothreshold (LVAT) features to support the regulatory approval of the AUTOGEN CRT-D family of devices.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Treatment
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Subjects prescribed a CRT-D and indicated per guidelines, who will receive an AUTOGEN CRT-D model G160, G161, G172, G173
- •Subjects who have an implanted functional endocardial right ventricular defibrillation lead or who will receive a endocardial right ventricular defibrillation lead
- •Subjects who have an implanted functional bipolar or unipolar left ventricular lead or who will receive a bipolar or unipolar left ventricular lead
- •Subjects who are willing and capable of providing informed consent to undergo a device implant procedure, and to participate in all testing and follow-ups defined in this protocol
- •Subjects whose age is 18 or above, or of legal age to give informed consent specific to national law
排除标准
- •Subjects who have an implanted multipolar (>2poles) left ventricular lead or who will receive a multipolar (>2poles) left ventricular lead
- •Subjects with an unknown model/manufacturer, or implant date for the RA, RV or LV lead
- •Subjects for whom a RV defibrillation lead manufactured by St. Jude Medical or Biotronik is implanted, is planned to be implanted, or has been abandoned
- •Implanted with an active Medtronic Sprint Fidelis® lead models: 6930, 6931, 6948 or 6949
- •Subjects with an implanted or abandoned St. Jude Medical QuickSite® or QuickFlex® lead models: 1056T, 1058T, 1156T, 1158T
- •Subjects with a RV or LV lead revision or extraction within 30 days of enrollment
- •Subjects with an implanted lead that is planned to be extracted during the study implant procedure
- •Subjects with an active implanted RA or RV lead that is greater than 10 years old, unless the lead will be abandoned
- •Subjects with an active implanted LV lead that is greater than 8 years old, unless the lead will be abandoned
- •Subjects preexisting unipolar pacemaker that will not be explanted/abandoned
- •Subjects with a life expectancy less than 6 months
- •Subjects with a prosthetic mechanical tricuspid heart valve
- •Women of childbearing age who are pregnant or plan to become pregnant. NOTE: women of childbearing potential with an uncertain pregnancy status must have a negative pregnancy test within 7 days prior to enrollment. Pregnancy tests are required as part of standard routing clinical practice at all centers whenever female patients are exposed to x-ray radiation. Since this protocol does not require any exposure of the female patients to x-ray and x-ray exposure is part of the clinical procedure, it is the responsibility of the investigators to ensure that pregnant females will not be exposed to x-ray at any time
- •Subject enrolled in a concurrent study, except national/governmental registries that do not require a signed informed consent form, without the written approval from Boston Scientific
- •Subjects who are not geographically stable, to the extent that it would prevent attending the study follow-ups at the investigational center
研究组 & 干预措施
CRT-D
For all subjects with the AUTOGEN CRT-D, the autothreshold algorithms will be evaluated
干预措施: CRT-D (Device)
结局指标
主要结局
Primary Safety Endpoint is to Evaluate the System-related Complication-free Rate
时间窗: 3 months
Safety of the AUTOGEN was evaluated by the system-related complication-free rate (CFR) at 3-months post-implant. The system consists of the implanted AUTOGEN CRT-D pulse generator, RA lead (if implanted), RV lead, and LV lead.
The Accuracy of the RVAT Commanded Test Will be Evaluated by Comparing the RVAT Determined Threshold to a Core Lab (Independent Physician) Determined Threshold.
时间窗: 1-month and 3-month follow-up visits
Accuracy of the algorithm will be measured for all patients, by comparing algorithm determined threshold to a core lab determined threshold at both the 1-month and 3-month follow-up visits. The RVAT determined threshold for all patients at 1-month and 3-month follow-up visits are pooled for final analysis. An accurate commanded threshold is defined by: \|commanded threshold - core lab determined threshold\| ≤ 0.2 V; if the commanded threshold is ≤ 3.5V or \|commanded threshold - core lab determined threshold\| ≤ 0.5 V; if the commanded threshold is \> 3.5V. Subjects were allowed to contribute multiple paired datasets for this endpoint analysis, one set each from the 1-month and 3-month visits. Paired datasets from the 1-month and 3-month visits were pooled for the purpose of this endpoint analysis.
The Accuracy of the LVAT Commanded Test Will be Evaluated by Comparing the LVAT Determined Threshold to a Core Lab (Independent Physician) Determined Threshold.
时间窗: 1-month and 3-month follow up visits
Accuracy of the algorithm will be measured for all patients, by comparing algorithm determined threshold to a core lab determined threshold at both the 1-month and 3-month follow-up visits. The LVAT determined threshold for all patients at 1-month and 3-month follow-up visits are pooled for final analysis. An accurate commanded threshold is defined by: commanded threshold - core lab determined threshold\| ≤ 0.2 V; if the commanded threshold is ≤ 3.5V or \|commanded threshold - core lab determined threshold\| ≤ 0.5 V; if the commanded threshold is \> 3.5V. Paired datasets from the 1-month and 3-month visits were pooled for the purpose of this endpoint analysis. Subjects were allowed to contribute multiple paired datasets for this endpoint analysis, one set each from the 1-month and 3-month visits. Paired datasets from the 1-month and 3-month visits were pooled for the purpose of this endpoint analysis.
The Accuracy of the RVAT Ambulatory Test Will be Evaluated by Comparing the RVAT Determined Threshold to a Core Lab (Independent Physician) Determined Threshold
时间窗: 1-month and 3-month follow up visits
Accuracy of the algorithm will be measured for all patients, by comparing algorithm determined threshold to a core lab determined threshold at both the 1-month and 3-month follow-up visits.The RVAT determined threshold for all patients at 1-month and 3-month follow-up visits are pooled for final analysis. An accurate Ambulatory threshold is defined by:\|Ambulatory threshold - ECG threshold\| ≤ 0.6 V; if the ECG threshold is ≤ 3.5V or \|Ambulatory threshold - ECG threshold\| ≤ 1.0 V; if the ECG threshold is \> 3.5V. Subjects were allowed to contribute multiple paired datasets for this endpoint analysis, one set each from the 1-month and 3-month visits. Paired datasets form the 1-month and 3-month visits were pooled for purposes of endpoint analysis.
The Accuracy of the LVAT Ambulatory Test Will be Evaluated by Comparing the LVAT Determined Threshold to a Core Lab (Independent Physician) Determined Threshold
时间窗: 1-month and 3-month follow up visits
Accuracy of the algorithm will be measured for all patients, by comparing algorithm determined threshold to a core lab determined threshold at both the 1-month and 3-month follow-up visits. The LVAT determined threshold for all patients at 1-month and 3-month follow-up visits are pooled for final analysis. An accurate Ambulatory threshold is defined by: \|Ambulatory threshold - ECG threshold\| ≤ 1.0 V. Paired datasets from the 1-month and 3-month visits were pooled for the purpose of this endpoint analysis. Subjects were allowed to contribute multiple paired datasets for this endpoint analysis, one set each from the 1-month and 3-month visits. Paired datasets form the 1-month and 3-month visits were pooled for purposes of endpoint analysis.
次要结局
- RVAT Secondary Efficacy Endpoint: The Percent of RVAT Commanded Tests That Result in an Appropriate Outcome(3-month follow up visit)
- LVAT Secondary Efficacy Endpoint: The Percent of LVAT Commanded Tests That Result in an Appropriate Outcome(3-month follow up visit)