MedPath

A Wearable In-phase Chest Wall Vibration Device for Relief of Dyspnoea in COPD: a First-in-human Exploratory Study

Not Applicable
Completed
Conditions
Chronic Obstructive Pulmonary Disease
Dyspnea
Interventions
Device: ResWave
Registration Number
NCT06432920
Lead Sponsor
Elevre Medical Ltd.
Brief Summary

The goal of this exploratory clinical trial is to evaluate the initial clinical safety and performance of a prototype wearable chest wall vibration (CWV) device intended to relieve exertional dyspnea in adults with Chronic Obstructive Pulmonary Disease. The main questions it aims to answer are:

* Can CWV be delivered safely to participants via a wearable device?

* Is there evidence of an effect on participant-reported dyspnea, endurance time and other measures of cardiorespiratory function?

Participants will undergo two cycle-ergometer exercise testing sessions while wearing the prototype wearable device. The device will be active in one session (intervention) and inactive in the other (control). The order in which intervention or control occurs will be randomised. Researchers gather data relating to adverse events, device deficiencies, participant-reported symptom severity, endurance time and measures of cardiorespiratory function recorded via standard CPET apparatus.

Detailed Description

The purpose of this Clinical Investigation is to evaluate the device design in a small number of human subjects with respect to initial clinical safety, and to gather preliminary data regarding the effect of the device on patient reported dyspnoea and physiological measures of cardiorespiratory function. This will facilitate planning of further steps of device development and validation, guide design modifications and define parameters for a future pivotal clinical investigation.

The investigational device is at Pilot stage of Clinical Development, and the proposed Clinical investigation is a first-in-human exploratory study. It is not a superiority, non-inferiority, or equivalence study.

The primary objective of this Clinical Investigation is to evaluate the investigational device with respect to initial clinical safety when used by subjects with Chronic Obstructive Pulmonary Disease during physical exertion in the form of cycle ergometer-based Cardiopulmonary Exercise Testing.

The secondary objective of this Clinical Investigation is to gather preliminary data regarding the effect of the device on patient-reported dyspnoea and physiological measures of cardiorespiratory function.

Each subject will attend the investigation site for a Pre-Study visit during which formal written consent will be obtained, clinical history and medications will be reviewed as part of standard CPET pre-assessment, and enrolment in the Clinical Investigation will be complete.

An initial Incremental Cardiopulmonary Exercise Testing (CPET) session will then be conducted to determine baseline Peak Work Rate capacity. This is necessary so that the intensity of CPET that will be required to achieve 75% of each subject's individual Peak Work Rate capacity can be determined. Peak Work Rate capacity will be defined as the highest work rate that the Subject is able to maintain for ≥30 seconds.

Following initial CPET, subjects will return for device testing on a non-consecutive day. Each subject will undergo 2 study arms during this visit: the Intervention (ResWave) and Control arms, in the random order. Both arms will be conducted on the same day, with a recovery period of 60 minutes between each arm. Subjects will undergo sub-maximal exercise testing at a Constant Work Rate (CWR) of 75% of their peak work rate capacity in both arms. Borg CR10 scores will be recorded at 2 minute intervals throughout. Endurance time will be recorded at test completion. Adverse Events will be recorded throughout testing and at 72-hour follow-up via phone call. Device Deficiencies will be recorded throughout testing.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
20
Inclusion Criteria
  • Diagnosis of Chronic Obstructive Pulmonary Disease (COPD) with FEV1 / FVC ratio of < 75% and deemed clinically stable
  • Moderate to severe airflow limitation on Spirometry (FEV1: 30 to 75% predicted)
  • Pulmonary Function Testing completed in the last 6 months
Exclusion Criteria
  • Active pulmonary infection or exacerbation within last 6 weeks
  • Unstable cardiac disease - cardiac intervention, change in symptoms or medication within last 6 weeks
  • BMI > 35
  • Receiving domiciliary oxygen therapy
  • Diagnosis of COVID-19 within the last 12 weeks or persistent symptoms directly attributable to previous COVID-19 infection
  • Presence of comorbid condition that is deemed to be a significant contributor to dyspnoea: Heart Failure, Chronic Thromboembolic Disease, Neuromuscular Disease, Primary Pulmonary Hypertension
  • Pulmonary Embolism in last 3 months
  • Abdominal or Thoracic surgery in last 3 months
  • Pneumothorax in last 6 months
  • Active malignancy
  • Active chest wall pain, active skin condition or open wound on thorax
  • Pregnant or breastfeeding
  • Women of child-bearing age (premenopausal, under age 50, or without previous sterilisation surgery)
  • Any implantable electronic device
  • On anticoagulants or with known history of coagulopathy
  • Any absolute or relative contraindication to CPET testing
  • Disability or comorbid condition that prevents exercise training and/or use of CPET apparatus and/or wearing the investigational device
  • Previous experience of Chest Wall Vibration therapies for relief of breathlessness
  • Currently enrolled in any other clinical trial or research study
  • People deemed to be incapable of giving consent, or with reduced capacity to consent or diminished autonomy as a result of mental or cognitive impairment, or deemed otherwise vulnerable on clinical grounds

Study & Design

Study Type
INTERVENTIONAL
Study Design
CROSSOVER
Arm && Interventions
GroupInterventionDescription
ResWaveResWaveIn-phase Chest Wall Vibration worn and active during constant work rate exercise testing.
ControlResWaveIn-phase Chest Wall Vibration worn and inactive (off) during constant work rate exercise testing.
Primary Outcome Measures
NameTimeMethod
Adverse Events72 hours

Adverse Events during or following use of the prototype device and/or exercise testing protocol

Device DeficienciesDuring active use in exercise testing

Performance data of the device during use

Secondary Outcome Measures
NameTimeMethod
Dyspnea severityRecorded at 2 minute intervals during exercise testing. Outcome measure at isotime.

Borg Category-Ratio scale (Borg CR10). Scale from 0 (lowest intensity, better) to 10 (maximal intensity, worse).

Endurance timeDuration of exercise testing

Exercise testing endurance time

Reason for stoppingDuring exercise testing

Whether participant stopped exercise testing due to dyspnea or other reason

Minute Ventilation (VE)During exercise testing

Minute Ventilation in millilitres/minute (mL/min)

Oxygen PulseDuring exercise testing

Oxygen Pulse (VO2/HR) in millilitres/minute (mL/min)

Heart Rate (HR)During exercise testing

Heart rate in beats per minute (BPM)

Oxygen ConsumptionDuring exercise testing

Oxygen Consumption (VO2) in millilitres/minute (mL/min)

Carbon Dioxide ProductionDuring exercise testing

Carbon Dioxide Production (VCO2) in millilitres/minute (mL/min)

WorkDuring exercise testing

Work (W) in watts

Tidal Volume (VT)During exercise testing

Tidal Volume in Litres (L) at body temperature and pressure saturated (BTPS)

Respiratory Quotient (RQ)During exercise testing

Respiratory Quotient (CO2 produced / O2 consumed; no unit as it is a ratio value)

Oxygen Saturation (SpO2)During exercise testing

Oxygen Saturation (SpO2) (%, percentage)

End-tidal carbon dioxide tension (PETCO2)During exercise testing

End-tidal carbon dioxide tension (PETCO2) in millimetres of mercury (mmHg)

End-tidal oxygen tension (PETO2)During exercise testing

End-tidal oxygen tension (PETO2) in millimetres of mercury (mmHg)

Respiratory ReserveDuring exercise testing

Respiratory Reserve in Litres

Expiration TimeDuring exercise testing

Expiration Time in seconds

Trial Locations

Locations (1)

Beacon Hospital

🇮🇪

Sandyford, Dublin 18, Ireland

© Copyright 2025. All Rights Reserved by MedPath