Comparison of an Optical System to Continuously Measure Arterial Blood Pressure Against a Gold-standard Arterial Catheter

CSEM Centre Suisse d'Electronique et de Microtechnique SA - Recherche et Developpement1 个研究点 分布在 1 个国家目标入组 70 人开始时间: 2016年4月1日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
70
试验地点
1
主要终点
Error (in mmHg) between paired determinations of arterial blood pressure from an optical device against reference measurements from an arterial catheter

研究概览

简要总结

Comparison of an optical method to continuously measure blood pressure against an invasive arterial catheter.

详细描述

The measurement of blood pressure during anesthesia is commonly performed by the inflation of a brachial cuff providing only intermittent blood pressure measurements. In some case, it is required to add a continuous invasive monitoring of blood pressure, via the insertion of an arterial catheter. Unfortunately, the related morbidity is not negligible. To limit the consequences of such a gesture, some developments have been done in the past to allow for continuous non-invasive measurements of blood pressure during anesthesia. Unfortunately, most of these devices are subject to important limitations and constraints of use.

A new system has been designed by CSEM consisting of an optical system fixed on patient's skin, absolutely non-invasive and easy to use. In order to study the reliability of this device, the investigators plan to compare its blood pressure estimates against gold-standard arterial catheter measurements during induction of general anesthesia on:

  • 40 patients wearing the optical system at the chest.
  • and 40 patients wearing the optical system at the fingertip

研究设计

研究类型
Interventional
分配方式
Non Randomized
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • patient requiring the placement of an arterial catheter for general anesthesia
  • minimum of 18 years
  • written consent

排除标准

  • minor patient or incapable of discernment
  • patient with AIMD
  • thoracic surgery

研究组 & 干预措施

OBPM 2015-MD-0022 - CHEST

Experimental

Cohort of 30 patients undergoing general anesthesia, monitored by optical signals at the chest

干预措施: Experimental: OBPM 2015-MD-0022 - CHEST (Device)

OBPM 2015-MD-0022 - CHEST & FINGER

Experimental

Cohort of 10 patients undergoing general anesthesia, monitored by optical signals at the chest and at the fingertip

干预措施: Experimental: OBPM 2015-MD-0022 - CHEST & FINGER (Device)

OBPM 2015-MD-0022 - FINGER

Experimental

Cohort of 30 patients undergoing general anesthesia, monitored by optical signals at the fingertip

干预措施: Experimental: OBPM 2015-MD-0022 - FINGER (Device)

结局指标

主要结局

Error (in mmHg) between paired determinations of arterial blood pressure from an optical device against reference measurements from an arterial catheter

时间窗: 24 months

The signals recorded by the optical blood pressure device will be retrospectively analysed (offline analysis) in order to test the hypothesis that blood pressure values as derived by the new device is comparable to blood pressure values as measured by an arterial catheter within the limits of agreement in mmHg stated by ISO 81060-2 2013.