Comparison of Diagnostic Accuracy of Automated Volume Breast Scanner to Hand Held Ultrasound for Diagnostic Breast Exam
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 90
- 主要终点
- Diagnostic Accuracy
研究概览
简要总结
Women age >18 scheduled for a diagnostic ultrasound examination had both a hand held ultrasound and an automated breast volume scan. Exams were performed by different technologists. Images were blinded and blinded read for diagnostic accuracy of the two examinations
详细描述
Women aged 18 years and above presenting for a diagnostic ultrasound secondary to a palpable mass, abnormal mammogram, follow-up ultrasound abnormality, or discharge were asked to participate in the study. If the patient had both a palpable mass and a mammogram that had an abnormality the indication was listed as a palpable abnormality. Exclusion criteria included inability to give informed consent and abnormality of the breast that is larger than one quadrant. No patients were excluded from the study.
All patients received both a Hand Held (HH) and Automated Breast Volume Scanner (ABVS) ultrasound. Patients were initially randomized to have either a HH or an ABVS. The HH ultrasound was performed by one of 4 sonographers that had a minimum of 10 years experience in breast ultrasound. A different sonographer than the one performing the HH or mammography technician performed ABVS. All technologists performing ABVS participated in a training course on ABVS and performed at least 10 patients satisfactorily before beginning the study. The choice of a mammography technician or sonography to perform the ABVS was randomized. Both the sonographer performing the HH and the sonographer or mammography technologist performing the ABVS were provided the patient's history and results of the mammogram if performed for placement of the ABVS and region of interest for the HH. However the results of the HH or ABVS were not available to the person performing the second examination.
Final diagnosis was performed either by a 12 gauge core biopsy (Celero, Hologic, Marlborough, MA ) or confirmation of a benign lesion by no significant change over 2 years.
研究设计
- 研究类型
- Observational
- 观察模型
- Case Only
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- Female
- 接受健康志愿者
- 否
入选标准
- •Women aged 18 years and above presenting for a diagnostic ultrasound secondary to a palpable mass, abnormal mammogram, follow-up ultrasound abnormality, or discharge were asked to participate in the study.
排除标准
- •Unable to give informed consent
- •breast abnormality greater than one quadrant
结局指标
主要结局
Diagnostic Accuracy
时间窗: one day
Diagnostic performance assuming BI-RADS category 1, 2, and 3 were benign and BI-RADS category 4 and 5 were malignant and using biopsy or clinical stability for two years as gold standard was assessed through ROC curves analysis. Sensitivity, specificity, Area Under the Curve (AUC) are presented along with 95% CI. Pairwise comparison of AUC between readers and methodologies are also given using z-statistics (6). P\<0.05 was considered statistically significant. All tests were two-sided.
次要结局
- Comparison of sonographer and mammography technologist performing Automated volume breast scanner(1 day)
研究者
Richard G. Barr MD, PhD
Medical Director
Northeastern Ohio Radiology Research and Education Fund