临床试验/NCT02388815
NCT02388815已完成不适用

Evaluation of the Accuracy of the Abbott Sensor Based Glucose Monitoring System - Paediatric Label Extension Study (CE)

Abbott Diabetes Care18 个研究点 分布在 1 个国家目标入组 89 人开始时间: 2015年3月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
89
试验地点
18
主要终点
Point Accuracy

研究概览

简要总结

To evaluate the point accuracy of the Abbott Sensor Based Glucose Monitoring System when used at home by children with diabetes.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
盲法
None

入排标准

年龄范围
4 Years 至 17 Years(Child)
性别
All
接受健康志愿者

入选标准

  • Age ≥ 4 years, ≤17 years
  • Each participant has an identified Caregiver of ≥18 years
  • Type 1 or type 2 diabetes using insulin administered by injections or continuous subcutaneous insulin infusion (CSII)
  • Currently testing blood glucose, on average at least 2 times per day
  • In the investigator's opinion, technically capable of using device (participant and/or caregiver).

排除标准

  • Concomitant disease or condition that may compromise patient safety including and not limited to; cystic fibrosis, severe mental illness, known or suspected eating disorder or any uncontrolled long term medical condition
  • Currently prescribed oral steroid therapy for any acute or chronic condition
  • Currently receiving dialysis treatment or planning to receive dialysis during the study
  • Female participant known to be pregnant
  • Participating in another study of a glucose monitoring device or drug that could affect glucose measurements or management
  • Currently using a continuous glucose monitoring (CGM) device
  • Known (or suspected) allergy to medical grade adhesives
  • In the investigator's opinion the participant or caregiver is unsuitable to participate due to any other cause/reason (both patient and caregiver considered)

结局指标

主要结局

Point Accuracy

时间窗: 14 days

Point accuracy of Sensor based glucose values versus fingerstick blood glucose determined as % within Consensus Error Grid zone A. The Consensus Error Grid was developed from a survey of 100 clinicians to evaluate the accuracy of glucose measurements. Glucose results from the system under test (y) are paired with those from a reference method (x) and each (x,y) point is plotted on a grid. The grid has 5 risk categories, assigned by the clinicians surveyed. Risk categories (in order of increasing severity) are: Zone A: no effect on clinical action; Zone B: altered clinical action or little or no effect on clinical outcome; Zone C: altered clinical action likely to effect clinical outcome; Zone D: altered clinical action, could have significant medical risk; Zone E: altered clinical action, could have dangerous consequences. Result were calculated for all subjects ie total number of sensor results and fingerstick blood glucose results divided by the total number of results x 100.

次要结局

未报告次要终点

研究者

申办方类型
Industry
责任方
Sponsor

研究点 (18)

Loading locations...