Validation of User Needs of the Point Digit With Partial Hand Amputees

Not Applicable
Completed
Conditions
Amputation; Traumatic, Hand
Interventions
Device: Point Digit
Registration Number
NCT04763278
Lead Sponsor
Point Designs
Brief Summary

The objective of the device feasibility study will be to validate the user needs of the Point Digit system. This study will be a single group intervention model where one group of 5 partial hand amputees will be asked to perform several tasks. Successful completion of a task results in a fulfilled user need. Failure to complete a task results in an unfulfilled user need.

Detailed Description

Not available

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
7
Inclusion Criteria
  • Partial hand amputees with absence of index and/or middle fingers and presence of thumb. Additional finger loss is acceptable if all other criterion are met.
  • Fluent in English
  • Individuals aged 18 or greater
Exclusion Criteria
  • Significant cognitive deficits as determined upon clinical evaluation
  • Significant neurological deficits as determined upon clinical evaluation
  • Significant physical deficits of the residual limb impacting full participation in the study as determined upon clinical evaluation
  • Uncontrolled pain or phantom pain impacting full participation in the study as determined upon occupational therapist evaluation
  • Serious uncontrolled medical problems as judged by the project therapist.

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
ProsthesisPoint DigitPatient is temporarily fit with Point Digit partial hand prosthetic system
Primary Outcome Measures
NameTimeMethod
VAL-1: System Comfort Validation TestDay 1 (8 hours)

User wears Point Digit(s) for a full day (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall be comfortable for a full day's use. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

VAL-2: Unilateral Function Validation TestDay 1 (10 minutes)

User flexes and extends the Point Digit(s) without using contralateral hand (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall allow the user to flex and extend finger without their second hand. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

VAL-3.1: System Robustness Validation Test With 25 lb BagDay 1 (10 minutes)

User lifts a 25 lb. bag with Point Digit(s) (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

VAL-3.2: Robustness Validation Test With HammerDay 1 (10 minutes)

User grips a hammer with Point Digit(s) and drives a nail into a piece of wood (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

VAL-4: Object Release Validation TestDay 1 (10 minutes)

User releases Point Digit(s) while performing an active grasp by pressing the dorsal button (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall be releasable while performing active grasps. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

VAL-5: Position Function Validation TestDay 1 (10 minutes)

User positions Point Digit(s) into each of the locking positions (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall have a smooth ratcheting mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

VAL-6: Extension Function Validation TestDay 1 (10 minutes)

User activates spring-back mechanism by using (1) contralateral hand, (2) tabletop, and (3) side or thigh (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall have a reliable auto spring-back mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

VAL-7: Grip Functions Validation TestDay 1 (10 minutes)

User grips a 5 lb. cylindrical smooth object using Point Digit(s) for 10 seconds (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall have high grip surface friction. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

VAL-8: Fine Motor Function Validation TestDay 1 (10 minutes)

User picks up 1 coin from a smooth tabletop (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall have a pronounced fingernail. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

VAL-9: Overall Function & Robustness Validation TestDay 1 (8 hours)

User performs VAL-1 through VAL-8 (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall be robustly attached to mounting system. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

Secondary Outcome Measures
NameTimeMethod

Trial Locations

Locations (2)

University of Colorado Denver | Anschutz Medical Campus

🇺🇸

Aurora, Colorado, United States

Point Designs

🇺🇸

Lafayette, Colorado, United States