Validation of User Needs of the Point Digit With Partial Hand Amputees
- Conditions
- Amputation; Traumatic, Hand
- Interventions
- Device: Point Digit
- Registration Number
- NCT04763278
- Lead Sponsor
- Point Designs
- Brief Summary
The objective of the device feasibility study will be to validate the user needs of the Point Digit system. This study will be a single group intervention model where one group of 5 partial hand amputees will be asked to perform several tasks. Successful completion of a task results in a fulfilled user need. Failure to complete a task results in an unfulfilled user need.
- Detailed Description
Not available
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 7
- Partial hand amputees with absence of index and/or middle fingers and presence of thumb. Additional finger loss is acceptable if all other criterion are met.
- Fluent in English
- Individuals aged 18 or greater
- Significant cognitive deficits as determined upon clinical evaluation
- Significant neurological deficits as determined upon clinical evaluation
- Significant physical deficits of the residual limb impacting full participation in the study as determined upon clinical evaluation
- Uncontrolled pain or phantom pain impacting full participation in the study as determined upon occupational therapist evaluation
- Serious uncontrolled medical problems as judged by the project therapist.
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Arm && Interventions
Group Intervention Description Prosthesis Point Digit Patient is temporarily fit with Point Digit partial hand prosthetic system
- Primary Outcome Measures
Name Time Method VAL-1: System Comfort Validation Test Day 1 (8 hours) User wears Point Digit(s) for a full day (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall be comfortable for a full day's use. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.
VAL-2: Unilateral Function Validation Test Day 1 (10 minutes) User flexes and extends the Point Digit(s) without using contralateral hand (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall allow the user to flex and extend finger without their second hand. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.
VAL-3.1: System Robustness Validation Test With 25 lb Bag Day 1 (10 minutes) User lifts a 25 lb. bag with Point Digit(s) (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.
VAL-3.2: Robustness Validation Test With Hammer Day 1 (10 minutes) User grips a hammer with Point Digit(s) and drives a nail into a piece of wood (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.
VAL-4: Object Release Validation Test Day 1 (10 minutes) User releases Point Digit(s) while performing an active grasp by pressing the dorsal button (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall be releasable while performing active grasps. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.
VAL-5: Position Function Validation Test Day 1 (10 minutes) User positions Point Digit(s) into each of the locking positions (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall have a smooth ratcheting mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.
VAL-6: Extension Function Validation Test Day 1 (10 minutes) User activates spring-back mechanism by using (1) contralateral hand, (2) tabletop, and (3) side or thigh (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall have a reliable auto spring-back mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.
VAL-7: Grip Functions Validation Test Day 1 (10 minutes) User grips a 5 lb. cylindrical smooth object using Point Digit(s) for 10 seconds (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall have high grip surface friction. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.
VAL-8: Fine Motor Function Validation Test Day 1 (10 minutes) User picks up 1 coin from a smooth tabletop (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall have a pronounced fingernail. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.
VAL-9: Overall Function & Robustness Validation Test Day 1 (8 hours) User performs VAL-1 through VAL-8 (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall be robustly attached to mounting system. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.
- Secondary Outcome Measures
Name Time Method