Pulse Oximeter Motion Study
- Conditions
- Hypoxia
- Interventions
- Other: Motion
- Registration Number
- NCT03156140
- Lead Sponsor
- University of California, San Francisco
- Brief Summary
Evaluate the performance of 4 pulse oximeters in 10 healthy adult volunteers with three motions compared to non-motion control at three arterial oxygen saturation target plateaus.
- Detailed Description
Evaluate the performance of 4 pulse oximeters (Masimo Radical-7, Nihon-Koden OxyPal Neo, Nellcor N-600, and Philips Intellivue MP5) in 10 healthy adult volunteers. Three motions will be evaluated: tapping, pseudorandom, and volunteer-generated rubbing, adjusted to produce photoplethsmogram disturbance similar to arterial pulsation amplitude. During motion, inspired gases will be adjusted to achieve stable target plateaus of arterial oxygen saturation (SaO2) at 75%, 88%, and 100%. Pulse oximeter readings (SpO2) will be compared with SaO2 from simultaneous arterial blood samples
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 10
healthy adult English speaking -
pediatric, cardiac or pulmonary condition. non-english speaking
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Arm && Interventions
Group Intervention Description motion Motion Right hand performs three different motion types
- Primary Outcome Measures
Name Time Method Pulse oximetry reading immediate
- Secondary Outcome Measures
Name Time Method Arterial Blood gas immediate