Perioperative Monitoring of Cardiac Surgery Patients With the Vitalstream Physiological Monitor
- Conditions
- Stroke VolumeCardiac Output, Low
- Interventions
- Device: Vitalstream
- Registration Number
- NCT05905536
- Lead Sponsor
- Wake Forest University Health Sciences
- Brief Summary
The purpose of this study is to validate cardiac output and stroke volume derived from Vitalstream against Gold Standard measurements obtained using thermodilution. The Vitalstream device is a continuous noninvasive physiological monitor (Caretaker Medical LLC, Charlottesville, Virginia, further referred to as "CTM") provides heart rate, continuous noninvasive blood pressure (BP), respiratory rate, stroke volume and cardiac output.
- Detailed Description
The device is currently undergoing IEC 60601 Medical Safety Standards testing and has already passed all FDA-required emission and discharge testing. The Vitalstream tracks central aortic BP via pulse analysis, specifically Pulse Decomposition Analysis ("PDA"), of the peripheral pulse at a distal site, typically finger. The device uses a low pressure \[30-40 mmHg\], pump-inflated, finger cuff that pneumatically couples arterial pulsations via a pressure line to a custom-designed piezo-electric pressure sensor for detection and analysis. Physiological data are communicated wirelessly to a tablet-based user interface via Bluetooth. The Vitalstream monitor is most beneficial as it is non-invasive, is a minimal risk device, and for this protocol will not be utilized to make treatment decisions for the study subject.
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 58
- Adult post-cardiac surgery needing cardiopulmonary bypass and use of the pulmonary artery catheter (PAC) as standard of care, (s/p coronary artery bypass graft (CABG), valve, heart transplant, major aortic & other vascular surgery, and a combination of CABG and valve procedures)
- patients (age > 18) admitted to the Intensive Care Unit (ICU) with a pulmonary artery catheter (continuous or intermittent thermodilution cardiac output) and an arterial catheter for Blood Pressure (BP) monitoring (standard of care)
- Patients without an appropriately positioned or functioning (PAC) pulmonary artery catheter admitted to the cardiac surgery ICU after surgery. Patients on left ventricular assist device (LVAD) support
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Arm && Interventions
Group Intervention Description Vitalstream Vitalstream Device placed on the subject preoperatively in the holding room and continued postoperatively into the Cardiac Intensive Care Unit (ICU).
- Primary Outcome Measures
Name Time Method Cardiac output values hour 24 Cardiac output values correlation of cardiac output between the PA catheter as assessed by the CCO and the VitalStream - the amount of blood pumped by the heart minute - Cardiac output is calculated by multiplying stroke volume with heart rate - Cardiac output in humans is generally 5-6 L/min in an at-rest to more than 35 L/min in elite athletes during exercise. HR is determined by signals from the sinoatrial node, which automatically depolarizes at an intrinsic rate of 60 to 100 times each minute - Low-output symptoms are caused by the inability of the heart to generate enough cardiac output. This leads to reduced blood flow to the brain and other vital organs
Blood pressure values hour 24 correlation of blood pressure continuously being monitored by the Vitalstream and the subjects indwelling arterial line
- Secondary Outcome Measures
Name Time Method
Trial Locations
- Locations (1)
Atrium Health Wake Forest Baptist
🇺🇸Winston-Salem, North Carolina, United States