跳至主要内容
临床试验/NCT01031914
NCT01031914
已完成
不适用

Evaluation of an Algorithm to Detect Sleep and Wake in Continuous Positive Airway Pressure Users Using Paced Breathing

Philips Respironics1 个研究点 分布在 1 个国家目标入组 36 人2009年10月

概览

阶段
不适用
干预措施
未指定
疾病 / 适应症
Sleep Apnea, Obstructive
发起方
Philips Respironics
入组人数
36
试验地点
1
主要终点
Sleep/Wake Algorithm
状态
已完成
最后更新
12年前

概览

简要总结

The purpose of this engineering trial is to develop and validate an algorithm that will deliver Paced Breathing as a ramp feature to obstructive sleep apnea (OSA) subjects using Continuous Positive Airway Pressure (CPAP) therapy. In this trial the investigators will be evaluating the algorithm's ability to correctly distinguish between sleep and wake.

注册库
clinicaltrials.gov
开始日期
2009年10月
结束日期
2010年9月
最后更新
12年前
研究类型
Interventional
研究设计
Single Group
性别
All

研究者

责任方
Sponsor

入排标准

入选标准

  • Age 21-70
  • Diagnosis of obstructive sleep apnea (OSA)
  • Current adherent CPAP user (has been using CPAP nightly for at least 2 weeks).
  • On CPAP pressures of 5-10cm.
  • Subjects wishing to complete a day or evening appointment need to have significant daytime sleepiness (Epworth Sleepiness Scale score of 8 or above)
  • Able and willing to provide written informed consent
  • English speaking

排除标准

  • Participation in another interventional research study within the last 30 days
  • Major controlled or uncontrolled medical condition such as congestive heart failure, neuromuscular disease, renal failure etc.
  • Inability to tolerate nasal CPAP mask due to problems breathing solely through their nose.
  • Chronic respiratory failure or insufficiency with suspected or known neuromuscular disease, moderate or severe chronic obstructive pulmonary disease (COPD), or any condition with an elevation of arterial carbon dioxide levels while awake (PaCo2≥55mmHg)
  • Severe oxygen desaturation on the polysomnography (PSG), i.e. Sa02 \< 70% for 10% of the study.
  • Surgery of the upper airway, nose, sinus or middle ear within the past 90 days
  • Currently using supplemental oxygen
  • Regular use of sleeping pills or stimulants (\> 3 nights a week)
  • Currently working night shift or rotating day/night shift
  • Drowsy Driving or near miss accident in the past 6 months

结局指标

主要结局

Sleep/Wake Algorithm

时间窗: The performance of the algorithm will be evaluated in real time while the subject is wearing the device during the sleep study, an average of 08 hours.

We tested the ability of the Sleep/Wake algorithm to identify sleep an wake periods with precision, as compared to standard polysonography (PSG) measures, which was used as the gold standard, i.e. we tested the accuracy of the algorithm. Accuracy was defined as the proportion of true results (both true positives and true negatives)in the population and it was assesed using as 2 X 2 table, i.e. accuracy = number of true positives + number of true negatives/ number of true positives + false positives + false negatives +true negatives. where True positive = the algorithm tested correctly identified sleep, False positive = the algorithm tested incorrectly identified sleep, True negative = the algorithm tested correctly rejected awake periods, and False negative = the algorithm tested incorrectly rejected awake periods.

研究点 (1)

Loading locations...