MedPath

LYNG21: Ably LYNG Clinical Demonstration in Operative Environment

Not Applicable
Conditions
Monitoring, Physiologic
Interventions
Device: LYNG by Ably Medical
Registration Number
NCT05117606
Lead Sponsor
Ably Medical AS
Brief Summary

The study main objective will be to compare measurements of heart rate, respiration rate, blood pressure and movement done by LYNG by Ably Medical with the corresponding measurements made by standard equipment in a patient sample in a hospital setting. The comparison will show the degree of agreement between LYNG and the existing current standard equipment. Data from the study will support Ably Medical in optimizing the LYNG product and measurement precision.

Detailed Description

The LYNG sensor mat is placed beneath a regular bed matrass and captures data without direct patient contact. Data collected by the LYNG sensor is transmitted via cables to the bed-side connectivity box for initial local analysis and storage. Data will be extracted and formatted for post-analysis in additional software for verification and validation purposes.

Specifically, the LYNG study objectives are as follows:

1. collect sensor data from LYNG Solution to optimize the LYNG data algorithms to further improve parameter estimations of patient heart rate, respiratory rate, blood pressure and movement indicators.

2. Compare parameter estimation by LYNG against current and relevant standard of care measurement methods

The study design is a within-subjects comparison in parameter estimation with Ably LYNG as comparator to standard of care. Patients will take part in 60-minute sessions where measurements are collected in accidence to standard procedure for the standard equipment while the LYNG system collects the same data non- intrusively. The data will not be part of any clinical decision making or have any impact of the patient care. The study will take place during Q2-Q3/2021 with estimated 8-12 weeks duration including logistics, setup, data collection, study closure.

Recruitment & Eligibility

Status
UNKNOWN
Sex
All
Target Recruitment
20
Inclusion Criteria
  • For patients:

    • Admitted to hospital,
    • Able to provide voluntary, informed consent.
Exclusion Criteria
  • For patients:

    • Health condition that endangers patient by participation due to pain, rehabilitation, and/or negative impact on treatment or care, as evaluated by medical personnel,
    • Pregnancy,
    • Significantly reduced mobility that implies substantial strain for moving in and out of bed, as evaluated by medical personnel
    • Unable to provide informed, voluntary consent personally or by legal guardian, * below 18 years old,
    • Patients with implantable devices such as pacemakers, baclofen pumps, insulin pumps, etc, or
    • Any way identified as 'vulnerable'

The nurse handing recruitment will explicitly check for inclusion/exclusion criteria alignment and document this in the study Enrolment Form. The routine will be to check for formal criteria and to describe the study in such a manner that the patient understands content, and make sure the patient understands before consenting

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
VitalsignsLYNG by Ably MedicalData from patients (heart rate, respiration, movement, blood pressure) is measured with traditional, standard hospital equipment in parallell with data from LYNG mat captured non-intrusively in the same session.
Primary Outcome Measures
NameTimeMethod
Absolute Relative Error of Heart Rate (beats per minute)60 minutes

absolute relative error = (reference measure/LYNG measure)/Reference

Absolute Relative Error of Respiration Rate (respirations per minute)60 minutes

absolute relative error = (reference measure/LYNG measure)/Reference

Movement60 minutes

Comparison of movement as part of patient instruction compared to indications of movement in movement sensors. Qualitative analysis.

absolute relative error of Blood Pressure = (reference measure/LYNG measure)/Reference60 minutes

absolute relative error = (reference measure/LYNG measure)/Reference