Comparative Study of Conventional 1.5 and 3.0T MR Images With Synthetically Reconstructed MR Images
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 117
- 试验地点
- 6
- 主要终点
- Mean Diagnostic Image Quality Difference Between Conventional Versus Synthetic MR Utilizing a 5 Point Likert Scale
研究概览
简要总结
This study is being conducted for regulatory submission of the GE Healthcare MAGnetic resonance Image Compilation (MAGiC), a point-of-care synthetic MR reconstruction software for GE 1.5T and 3.0T MR scanners.
详细描述
This study is a prospective, blinded, multi-center multi-reader (MCMR) clinical trial that is statistically powered for non-inferiority of diagnostic image quality of synthetic MR versus conventional MR and collection of supporting radiologic findings and morphology data from radiologists that reflect anticipated clinical usage of synthetic MR technology.
The study consists of three parts:
- Clinical MR acquisition,
- synthetic MR post-processing at a GE facility, and
- blinded image evaluation (BIE) of conventional and synthetic images for each subject.
The study will enroll eligible adult subjects with clinical indications for brain MRI that meet applicable site MR safety criteria and are not pregnant. The study consists of one visit, the subject's MRI scan, without additional follow-up. Results are expected to be representative of expected clinical usage of synthetic MAGiC MR technology in clinical practice.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Diagnostic
- 盲法
- None
盲法说明
None used for this single group study.
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Subjects must:
- •Be an adult 18 years of age or older;
- •Have clinical indication for MRI of the brain according to the site standard of care;
- •If female of childbearing potential (not surgically sterile or post-menopausal), be determined non-pregnant by a medically qualified investigator or demonstrated non-pregnant by negative urine pregnancy test;
- •Be able to hear and understand instructions without assistive devices;
- •Have necessary mental capacity to understand instructions be able to comply with protocol requirements;
- •Are able and willing to provide written informed consent by signing the informed consent form (ICF).
排除标准
- •Subjects will be excluded that:
- •Were previously enrolled in this evaluation;
- •Have metallic/conductive or electrically/magnetically active implants or attached medical devices (except for dental devices/fillings, surgical clips, and surgical staples) that could be unsafe for MRI;
- •Have contraindication(s) to MRI scanning per the routine MR Safety Screening policy of the investigational site;
- •Have severe trauma or pre-existing pathology that is expected to interfere with normal conduct of MR scanning or complete scanning of the brain;
- •Have medical condition(s) such as those requiring urgent medical care that, in the opinion of a physician Investigator, would prevent safe participation in the study.
研究组 & 干预措施
Neurological MRI
Images acquired for post processing with synthetic software. This is a crossover design where all subjects receive the same imaging scan, and comparison is done between the raw conventional scan and post-processed images using the research software.
干预措施: Neurological MRI (Device)
结局指标
主要结局
Mean Diagnostic Image Quality Difference Between Conventional Versus Synthetic MR Utilizing a 5 Point Likert Scale
时间窗: 1 day
Images were read by board-certified radiologists who reviewed each image and assessed it on a 5 Point scale (1=unacceptable, 2=poor image quality, 3=Acceptable, 4=Good, 5=Excellent) based on the quality of the image (eg, signal-to-noise, clarity of anatomic boundaries, and other parameters). The outcome was reported as the mean diagnostic image quality score difference (synthetic - conventional).
次要结局
未报告次要终点