Clinical Trials/NCT02942836
NCT02942836
Completed
Not Applicable

CareTaker Respiratory Rate Detection Study

CareTaker Medical LLC0 sites21 target enrollmentMarch 2016

Overview

Phase
Not Applicable
Intervention
Not specified
Conditions
Respiration Rate Detection
Sponsor
CareTaker Medical LLC
Enrollment
21
Primary Endpoint
Respiration rate detection with accuracy of better than 3 breaths per minute
Status
Completed
Last Updated
9 years ago

Overview

Brief Summary

The goal of the proposed study is to test the capability of the CareTaker monitor to extract the respiration signature from its acquired signal, and to determine a respiration rate of breaths/minute that meets or exceeds the performance levels of its predicate device.

Detailed Description

The goal of the proposed study is to test the capability of the CareTaker monitor to extract the respiration signature from its acquired signal, and to determine a respiration rate of breaths/minute that meets or exceeds the performance levels of its predicate device. The detection approach of RSPR (respiratory rate) that the CareTaker technology takes is based on peak detection and pulse analysis of arterial pressure pulse signals, as well as spectral analysis of the raw CareTaker sensor signal. The study was performed under the supervision of the Chesapeake IRB and involved 21 subjects \>18 y who were monitored during spontaneous and metronome-paced breathing.

Registry
clinicaltrials.gov
Start Date
March 2016
End Date
June 2016
Last Updated
9 years ago
Study Type
Observational
Sex
All

Investigators

Responsible Party
Sponsor

Eligibility Criteria

Inclusion Criteria

  • \> 18 years of age, who are able and willing to participate and have given verbal assent

Exclusion Criteria

  • Unable to give written consent
  • \<18 years of age
  • No or poor finger pulse, as determined through visual inspection for ischemic hands

Outcomes

Primary Outcomes

Respiration rate detection with accuracy of better than 3 breaths per minute

Time Frame: 15 minutes of matched respiration rates

Comparison with performance of predicate device (accuracy within 3 breaths per minute).