Skip to main content
Clinical Trials/NCT02063841
NCT02063841CompletedEarly Phase 1

A Prospective Double-blind, Randomized, Sham-controlled Trial to Evaluate the Effect of a Lead Free Drape on Scatter Radiation Exposure to Endoscopy Staff During Endoscopic Retrograde Cholangiopancreatography

Yale University2 sites in 1 country2 target enrollmentStarted: August 2013Last updated:
Conditions

Trial Snapshot

Phase
Early Phase 1
Status
Completed
Enrollment
2
Locations
2
Primary Endpoint
Amount of radiation exposure

Study Overview

Brief Summary

ERCP is associated with radiation exposure to the endoscopist and staff, which may be significant at high volume programs despite the use of lead aprons. We hypothesize that draping of the fluoroscopy image intensifier may significantly reduce staff radiation exposure and help achieve implementation of the ALARA (As Low As Reasonably Achievable) principle.

Detailed Description

A total of 100 consecutive ERCP procedures were randomly assigned to the radiation-attenuating or sham drape in a 1:1 ratio. There were 2 subjects, the endoscopists performing ERCP and the nurse assisting in the procedure.

Study Design

Study Type
Interventional
Allocation
Randomized
Intervention Model
Crossover
Primary Purpose
Prevention
Masking
Double (Participant, Investigator)

Eligibility Criteria

Ages
45 Years to 50 Years (Adult)
Sex
All
Accepts Healthy Volunteers
No

Inclusion Criteria

  • All the ERCP procedures done by the same endoscopist and nurse are enrolled
  • Patients age >19

Exclusion Criteria

  • Fluoroscopic procedures other than ERCP were excluded.

Outcomes

Primary Outcomes

Amount of radiation exposure

Time Frame: Length of Endoscopic retrograde cholangiopancreatography, average 5 minutes

Comparing the use of a lead-free protective drape containing bismuth and antimony and a sterile sham protective drape, radiation dose (RD) was measured at 3 different sites (endoscopist's eye (1) and thyroid (2), and assisting nurse's thyroid (3)) with personal dosimeters placed outside of eyeglass frames and the thyroid shield. The unit of measurement is each event. The subject was subjected to radiation during 100 Endoscopic retrograde cholangiopancreatography procedures.Both groups used standard of care lead protection.

Secondary Outcomes

  • Estimate cumulative annual radiation risk(Length of Endoscopic retrograde cholangiopancreatography, average 5 minutes)

Investigators

Sponsor Class
Other
Responsible Party
Sponsor

Study Sites (2)

Loading locations...