Beyond Confounders: Addressing Source of Measurement Variability and Error in Shear Wave Elastography
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 30
- 试验地点
- 1
- 主要终点
- Deviations From Suggested Guidelines- Any Effect on Agreement
研究概览
简要总结
Chronic liver disease is a major problem in the general population and there is an unmet need to diagnose(and screen) for liver disease with using noninvasive, cost-effective and sensitive techniques.The investigators hypothesize that variation using ultrasound elastography for the estimation of stage of liver fibrosis and steatosis in patients with diffuse liver disease exists due to different methods of measurements, and/or different systems. The proposed investigation is a cross-sectional study using ultrasound elastography and fat quantification modalities. The investigators are planning to enroll 30 subjects 18 years old and older in whom diffuse liver disease is suspected, and who have undergone non-focal liver biopsy in the past 6 months or are scheduled to undergo biopsy within 3 months of enrollment, as part of their routine clinical care. The investigators will use 4 different ultrasound devices with their shear wave elastography and speed of sound functions.
Specific aims;
- Compare shear wave elastography(SWE) measurements from different ultrasound systems; using histopathology as reference standards.
- Assess intra-operator and inter-operator reliability by measuring variability in elastography values by two operators on a single system.
- Determine the effect of deviations from guidelines(less number of measurements and measurements during active breath)
详细描述
Liver disease and cirrhosis are important causes of morbidity and mortality in the United States and are a major public health problem with 40,000 deaths and more than 1.4 billion dollars spent on medical services. Hepatic fibrosis is the final common pathway for many different liver insults and is known to be a dynamic process, which is reversible if diagnosed early and treated. If untreated, fibrosis eventually progresses to cirrhosis, which is irreversible. The diagnosis of fibrosis relies on liver biopsy (the gold standard) however, is an invasive procedure with risks and sampling errors. Indirect biomarkers of fibrogenesis can measure some of these components to determine categories of fibrosis, with a sensitivity and specificity of 47% and 90% respectively.It is shown in the literature that, fat content of liver is related with fibrosis development. Biopsy is accepted as the most accurate technique to assess liver fat and fibrosis amount.
Fibrotic livers demonstrate increased stiffness, a property that can be measured using technology named Ultrasound Elastography or sonoelastography (SWE). SWE is performed by insonating the patient with a low energy, amplitude, and frequency shear wave created by a vibrating probe. The propagated wave travels faster with increasing fibrosis: the stiffer the tissue, the faster the shear wave propagates. A pulse-echo ultrasound acquisition allows measurement of the wave velocity and the results are presented as kilopascals (kPa). Prior reports have described sensitivity and specificity for liver fibrosis detection of 80% and 97% respectively for SWE. The benefits of SWE are that it is inexpensive, reproducible, painless, rapid (< 10 min), easy to perform, and can be used for diagnosis, prognosis and monitoring disease progression.
The objective of this study is to compare the variation in elastography values and study the factors that might cause these variations.
The proposed investigation is a cross-sectional study using ultrasound elastography. The investigators are planning to enroll 30 subjects 18 years old and older in whom diffuse liver disease is suspected, and who have undergone non-focal liver biopsy in the past 6 months or are scheduled to undergo biopsy within 3 months of enrollment, as part of their routine clinical care.
All subjects will be required to come to the MGH main campus for 2 study visits within 60 days of each other. In addition, subjects will need to fast for at least 4 hours prior to study visits
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Diagnostic
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Adult Patients
- •Men or Woman
- •Suspected diffuse liver disease and have had a liver biopsy within the last 6 months or are scheduled for a liver biopsy in the next 3 months.
- •Consent to participate in the study
排除标准
- •Pregnancy
- •Acute illness/ cognitive impairment resulting in inability to cooperate with ultrasound
- •Patients that do not consent to ultrasound examination
研究组 & 干预措施
30 Patients with known liver biopsy results
Patients with chronic liver disease with known biopsy results
干预措施: Sonographic SWE measurements with 4 different ultrasound systems (Device)
结局指标
主要结局
Deviations From Suggested Guidelines- Any Effect on Agreement
时间窗: Visit 1 and Visit 2, an average of 2.5 hours for each visit
We collected guideline suggested shear wave elastography measurements from 20 patients. Guidelines suggest collecting 1) 10 measurements, 2) Asking patient to hold breath during the measurements. As a deviation from guideline suggestions, we collected measurements during free breath movements without asking patient to hold breath. As a deviation from the guideline suggestions, fewer number of measurements (3 measurements) were collected with asking the patient to hold breath during the measurements. All measurements were collected using one type of ultrasound system. All measurements were collected by 2 operators in 2 visits.
Interoperator and Intraoperator Agreement in m/s and kPA Units
时间窗: Visit 1 and Visit 2, an average of 2.5 hours for each visit
In 20 patients we collected 10 shear wave elastography measurements in m/s and kPA units. Using one of the available ultrasound systems, 2 operators collected these measurements in 2 visits. Theoretically, m/s to kPA conversion can be made using an equation. However, some systems provide opportunity to get data in both units. Although algebraically m/s and kPA can be converted to each other, we found some minor differences in terms of inter-operator and intra-operator agreement.
次要结局
未报告次要终点
研究者
Anthony Samir
Service Chief,Body Ultrasound Department of Radiology, Assistant Professor at Harvard Medical School
Massachusetts General Hospital