Intraocular pressure comparisons between supine and sitting position by portable tonometers, and the evaluation the effect of ocular biometry.
Not Applicable
- Conditions
- Healty volunteers
- Registration Number
- JPRN-UMIN000008497
- Lead Sponsor
- Saneikai Tsukazaki Hospital
- Brief Summary
The repeatability of Icare PRO was slightly lower in the supine position than in the sitting position. Although Icare PRO underestimated IOP values in eyes with higher IOP when compared with Tonopen XL and Kowa HAT in both positions, we observed good inter-device agreement between Icare PRO and both Tonopen XL and Kowa HAT.
- Detailed Description
Not available
Recruitment & Eligibility
- Status
- Complete: follow-up complete
- Sex
- All
- Target Recruitment
- 120
Inclusion Criteria
Not provided
Exclusion Criteria
Subjects with any ocular diseases and surgeries
Study & Design
- Study Type
- Observational
- Study Design
- Not specified
- Primary Outcome Measures
Name Time Method Intraocular pressure
- Secondary Outcome Measures
Name Time Method