Intraocular pressure comparisons between supine and sitting position by portable tonometers, and the evaluation the effect of ocular biometry. - Supine and sitting IOP measurements with ocualr biometry

Saneikai Tsukazaki Hospital0 个研究点目标入组 120 人开始时间: 2012年7月23日最近更新:
适应症

试验速览

阶段
未知
状态
已完成
发起方
入组人数
120

研究概览

简要总结

The repeatability of Icare PRO was slightly lower in the supine position than in the sitting position. Although Icare PRO underestimated IOP values in eyes with higher IOP when compared with Tonopen XL and Kowa HAT in both positions, we observed good inter-device agreement between Icare PRO and both Tonopen XL and Kowa HAT.