SpO2 Measurement Depending on Different Pulse Oximeter Averaging Time Settings
- Conditions
- Reduced Perfusion During SpO2 MeasurementsMovement Artifacts During SpO2 Measurements
- Interventions
- Other: Movement artifact - knockingOther: Movement artifact - tremorOther: Artifact - ambient lightOther: Artifact - low perfusionOther: SpO2 measurement
- Registration Number
- NCT04933773
- Lead Sponsor
- Czech Technical University in Prague
- Brief Summary
The aim of this study is to characterize the noise in the SpO2 signal depending on different pulse oximeter averaging time settings.
- Detailed Description
Noise in the SpO2 signal is most often caused by artifacts, such as movement artifacts, reduced perfusion, additional light and others. These artifacts can be partially removed from the SpO2 signal by different pulse oximeter settings, specifically the sensitivity mode and averaging time. It is known that when a longer averaging time is set, the measured signal contains fewer artifacts, on the other hand there is an increased risk of missing short desaturation. The aim of this work is to characterize the noise in the SpO2 signal depending on the set averaging time of the pulse oximeter based on experimental data from healthy volunteers.
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 17
- healthy volunteers from the Czech Technical University
- pregnancy
- severe cardiovascular conditions
- injury to the upper limbs or hands that could affect the peripheral perfusion
- diabetes
- hypotension or hypertension
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- CROSSOVER
- Arm && Interventions
Group Intervention Description Set averaging times: 2-4 seconds and 8 seconds Artifact - ambient light SpO2 is measured by two pulse oximeters on one participant's hand (on the index finger and middle finger). One pulse oximeter is set to an averaging time of 2-4 seconds, the second pulse oximeter is set to an averaging time of 8 seconds. Set averaging times: 8 seconds and 16 seconds Movement artifact - tremor SpO2 is measured by two pulse oximeters on one participant's hand (on the index finger and middle finger). One pulse oximeter is set to an averaging time of 8 seconds, the second pulse oximeter is set to an averaging time of 16 seconds. Set averaging times: 8 seconds and 16 seconds SpO2 measurement SpO2 is measured by two pulse oximeters on one participant's hand (on the index finger and middle finger). One pulse oximeter is set to an averaging time of 8 seconds, the second pulse oximeter is set to an averaging time of 16 seconds. Set averaging times: 2-4 seconds and 8 seconds SpO2 measurement SpO2 is measured by two pulse oximeters on one participant's hand (on the index finger and middle finger). One pulse oximeter is set to an averaging time of 2-4 seconds, the second pulse oximeter is set to an averaging time of 8 seconds. Set averaging times: 2-4 seconds and 8 seconds Movement artifact - tremor SpO2 is measured by two pulse oximeters on one participant's hand (on the index finger and middle finger). One pulse oximeter is set to an averaging time of 2-4 seconds, the second pulse oximeter is set to an averaging time of 8 seconds. Set averaging times: 8 seconds and 16 seconds Artifact - low perfusion SpO2 is measured by two pulse oximeters on one participant's hand (on the index finger and middle finger). One pulse oximeter is set to an averaging time of 8 seconds, the second pulse oximeter is set to an averaging time of 16 seconds. Set averaging times: 2-4 seconds and 8 seconds Movement artifact - knocking SpO2 is measured by two pulse oximeters on one participant's hand (on the index finger and middle finger). One pulse oximeter is set to an averaging time of 2-4 seconds, the second pulse oximeter is set to an averaging time of 8 seconds. Set averaging times: 2-4 seconds and 8 seconds Artifact - low perfusion SpO2 is measured by two pulse oximeters on one participant's hand (on the index finger and middle finger). One pulse oximeter is set to an averaging time of 2-4 seconds, the second pulse oximeter is set to an averaging time of 8 seconds. Set averaging times: 8 seconds and 16 seconds Movement artifact - knocking SpO2 is measured by two pulse oximeters on one participant's hand (on the index finger and middle finger). One pulse oximeter is set to an averaging time of 8 seconds, the second pulse oximeter is set to an averaging time of 16 seconds. Set averaging times: 8 seconds and 16 seconds Artifact - ambient light SpO2 is measured by two pulse oximeters on one participant's hand (on the index finger and middle finger). One pulse oximeter is set to an averaging time of 8 seconds, the second pulse oximeter is set to an averaging time of 16 seconds.
- Primary Outcome Measures
Name Time Method SpO2 readings with different averaging times 30 minutes Differences in the SpO2 signal during artifacts depending on the set averaging time.
- Secondary Outcome Measures
Name Time Method
Trial Locations
- Locations (1)
Faculty of Biomedical Engineering, Czech Technical University in Prague
🇨🇿Kladno, Czechia