Skip to main content
Clinical Trials/NCT03495297
NCT03495297CompletedNot Applicable

A Prospective Randomised CompArative Trial of SubcutanEous ImplanTable CardiOverter-DefibrillatoR ImplANtation With and Without DeFibrillation Testing

Academisch Medisch Centrum - Universiteit van Amsterdam (AMC-UvA)32 sites in 4 countries965 target enrollmentStarted: May 7, 2018Last updated:
Conditions
Interventions

Trial Snapshot

Phase
Not Applicable
Status
Completed
Enrollment
965
Locations
32
Primary Endpoint
Failed first appropriate shock in a spontaneous episode

Study Overview

Brief Summary

This study will test the hypothesis that implantation of a subcutaneous implantable cardioverter defibrillator (S-ICD) without performing a defibrillation test is non-inferior to S-ICD implant with a defibrillation test with regards to the primary endpoint failed first shock in a spontaneous arrhythmia episode when implant position is confirmed with PRAETORIAN score.

Detailed Description

Implantable Cardioverter Defibrillator (ICD) implant improves survival in patients with a higher risk for sudden cardiac death. There are 2 types of ICD available; transvenous ICD (TV-ICD) and subcutaneous ICD (S-ICD). During ICD implant, defibrillation testing (DFT) is performed to test functionality of the device. However, DFT can be associated with complications such as inability to convert, complications arising from general anaesthesia, prolonged resuscitation, stroke and death. Whereas DFT may be associated with complications, the benefit of DFT is debated as literature shows there is only a modest average effect of DFT, if any, on mortality, shock efficacy or safety. Recently it was shown in a randomized controlled trial called 'SIMPLE' that routine defibrillation testing of TV-ICDs at the time of implant does not improve shock efficacy or reduce arrhythmic death. For S-ICD there is only limited data available of the effect of DFT on S-ICD efficacy. Data have however shown that the conversion efficacy of the S-ICD is comparable to TV-ICD.

DFT is currently performed in standard S-ICD implants, but is omitted in specific cases. However, an alternative method to evaluate the correct position may be desired when omitting DFT. The PRAETORIAN Score is developed using computer modelling data on factors influencing defibrillation thresholds. The PRAETORIAN score represents the chance of a patient having an elevated defibrillation threshold and consequently failing a DFT or conversion of a spontaneous arrhythmia episode. The score was retrospectively validated in two studies with 180 and 321 patients.

It was reported that most S-ICD implants are performed under general anesthesia, however other anesthesia protocols are used as well. One of the most predominant factors to use general anesthesia is the performance of the DFT. If this is omitted, other anesthesia protocols may be a good option for many patients as well.

Study Design

Study Type
Interventional
Allocation
Randomized
Intervention Model
Parallel
Primary Purpose
Supportive Care
Masking
None

Eligibility Criteria

Ages
18 Years to — (Adult, Older Adult)
Sex
All
Accepts Healthy Volunteers
No

Inclusion Criteria

  • •Patients who meet current guidelines for ICD therapy and intent to undergo a de novo implant procedure for an S-ICD
  • •Patients must pass S-ICD screening per local routine
  • •Patients over 18 years of age, willing and capable to give informed consent
  • •Patients must be willing and capable of complying to follow up visits
  • •Patient must be eligible for either DFT strategy per physician discretion

Exclusion Criteria

  • •Patient with a life expectancy shorter than 12 months due to any medical condition
  • •Patients known to be pregnant
  • •Patients with intracardiac thrombus
  • •Patients with atrial fibrillation without appropriate anticoagulation
  • •Patients likely to undergo heart transplant within 12 months
  • •Patients with LVAD
  • •Patients with other contra-indications for DFT per physician's discretion

Arms & Interventions

S-ICD Implant without defibrillation test

Experimental

Patients undergoing de novo S-ICD implantation without induction of VF and defibrillation testing post-implant

Intervention: ommitence of defibrillation testing (Procedure)

S-ICD Implant with defibrillation test

No Intervention

Patients undergoing de novo S-ICD implantation including induction of VF and defibrillation testing post-implant

Outcomes

Primary Outcomes

Failed first appropriate shock in a spontaneous episode

Time Frame: 40 months

Number of patients having experiences a failed first appropriate shock in a spontaneous arrhythmia episode

Secondary Outcomes

  • Conversion efficacy within 5 shocks in spontaneous episodes(40 months)
  • Pain post implant(1-4 hours post implant)
  • Time to therapy(40 months)
  • Time to succesful therapy(40 months)
  • MACE post DFT(30 days)
  • Cardiac (pre-)syncope(40 months)
  • Device or lead repositioning(40 months)
  • ICD related infection(40 months)
  • Composite complications 30 days after implant(30 days)
  • PRAETORIAN Score(up to 24 hours)
  • Overall DFT conversion success(40 months)
  • Length of hospitalization(40 months)
  • Appropriate ICD therapy(40 months)
  • Successful DFT(40 months)
  • Shock efficacy(40 months)
  • S-ICD related complications(40 months)
  • Mortality(40 months)
  • DFT related complications(24 hours or 30 days)
  • Inappropriate ICD therapy(40 months)
  • Cardiac decompensation(40 months)

Investigators

Sponsor Class
Other
Responsible Party
Principal Investigator
Principal Investigator

R.E. Knops

Principal Investigator

Academisch Medisch Centrum - Universiteit van Amsterdam (AMC-UvA)

Study Sites (32)

Loading locations...

Similar Trials