MightySat - Clinical Performance Comparison Study

Masimo Corporation1 个研究点 分布在 1 个国家目标入组 34 人开始时间: 2017年7月10日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
34
试验地点
1
主要终点
Accuracy of RRp Determination

研究概览

简要总结

This study compares the performance of respiratory rate from pleth (RRp) measured with MightySat device on healthy subjects.Subjects will be asked to breathe at specific rates ranging from 5 to 30 breaths per minute.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Other
盲法
None

入排标准

年龄范围
18 Years 至 45 Years(Adult)
性别
All
接受健康志愿者
是

入选标准

  • •18 to 45 years old
  • •Physical status of ASA I or II
  • •Must be able to read and communicate in English
  • •Has signed all necessary related documents, e.g. written informed consent, confidentiality agreement
  • •Passed health assessment screening
  • •Negative pregnancy test for female subjects of child bearing potential

排除标准

  • •Younger than 18 years old, older than 45 years old
  • •ASA physical status of III, IV, and V
  • •Subject has any medical condition which in the judgement of the investigator, renders them inappropriate for participation in the study
  • •Inability to tolerate sitting still or minimal movement for at least 30 minutes
  • •Positive pregnancy test for female subjects
  • •Refusal to take pregnancy test for women of child bearing potential
  • •Nursing female subjects
  • •Refusal of male subjects to agree to shave hair off areas where sensors will be applied (neck)
  • •Excluded at Principal Investigator's discretion

研究组 & 干预措施

MightySat Test group

Experimental

The subjects will be enrolled into the test group and will receive the MightySat investigational pulse oximeter.

干预措施: MightySat (Device)

结局指标

主要结局

Accuracy of RRp Determination

时间窗: approximately 30-40 minutes

Accuracy will be determined by comparing the noninvasive respiratory rate by pleth (RRp) measurement of the MightySat to the respiratory rate (RR) reference and calculating the arithmetic root mean square error (Arms) value. In order to obtain the Arms value, the RR measurement from the reference is subtracted from the MightySat RRp measurement for a number of samples, the average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the Arms Error value.

次要结局

未报告次要终点

研究者

申办方类型
Industry
责任方
Sponsor