Skip to main content
Clinical Trials/NCT02048514
NCT02048514CompletedNot Applicable

The Nellix® EndoVascular Aneurysm Sealing System for the Treatment of Infrarenal Abdominal Aortic Aneurysms

Endologix7 sites in 4 countries69 target enrollmentStarted: December 2008Last updated:
Conditions

Trial Snapshot

Phase
Not Applicable
Status
Completed
Sponsor
Enrollment
69
Locations
7
Primary Endpoint
Safety- Number of Major Adverse Events

Study Overview

Brief Summary

The purpose of this clinical study is to evaluate the safety and device performance of the Nellix® EndoVascular Aneurysm Sealing System (Nellix System) for the treatment of infrarenal abdominal aortic aneurysms.

Detailed Description

This is a multicenter, multi-country prospective, single-arm, non-randomized clinical trial. The comparative control arm is a combination of patient data drawn from the abdominal aortic aneurysms (AAA) open surgical control arms of four commercially approved AAA devices as part of the Society of Vascular Surgeons. Patients enrolled in this study will be pre-screened and have a diagnosis of an infrarenal AAA with or without the need for treatment of iliac artery aneurysm, are candidates for endovascular repair, and meet the eligibility criteria. Patients enrolled in the study will participate in follow-up visits 31 days, 6 months and at 1 year following the index procedure. At the follow-up visits there will be a clinical assessment, blood collection, and abdominal CT imaging of the patient.

Study Design

Study Type
Interventional
Allocation
Na
Intervention Model
Single Group
Primary Purpose
Treatment
Masking
None

Eligibility Criteria

Ages
18 Years to — (Adult, Older Adult)
Sex
All
Accepts Healthy Volunteers
No

Inclusion Criteria

  • At least 18 years of age
  • Informed consent form understood and signed and patient agrees to all follow-up visits
  • Is able and willing to comply with clinical follow-up requirements for one year
  • Is able and willing to undergo Contrast-Enhanced Spiral CT scans
  • Is a candidate for endovascular repair of an infrarenal aortic aneurysm with or without an iliac aneurysm with at least one of the following;
  • aneurysm ≥ 4.5 cm in diameter, or
  • aneurysm is twice the diameter of the normal infrarenal aorta, or
  • aneurysm is growing at a rate of ≥ 10 % per year
  • Non-aneurysmal aortic neck length ≥ 5mm below the most inferior renal artery
  • Proximal aortic neck diameter between 16 and 36mm
  • Aortic neck angulation to the sac ≤ 60
  • Common iliac artery diameter 8 to 35mm bilaterally
  • Aneurysm blood lumen diameter ≤ 60mm
  • Iliac and femoral arteries suitable for endovascular access with the Nellix System

Exclusion Criteria

  • Life expectancy ≤ 1 year
  • Dissecting, ruptured, or leaking aneurysm (blood outside the aorta), has sustained recent trauma, is symptomatic, or is non-degenerative
  • Aneurysm is thoraco-abdominal, suprarenal and/or mycotic
  • Thoracic aneurysm ≥ 4.5cm in diameter
  • AAA requires treatment that would result in bilateral occlusion of the internal iliac arteries
  • Presence of mural thrombus > 50% circumferentially in aortic neck
  • Iliac aneurysm blood lumen diameter > 35mm
  • Active infection, vasculitis, or systemic connective tissue disease (e.g., Rheumatoid arthritis, scleroderma, Marfan's syndrome, Ehlers-Danlos disease) at time of procedure
  • History of prior treatment of abdominal aortic or iliac artery aneurysm disease
  • Femoral or iliac artery occlusive disease such that device delivery is not possible
  • Vascular access for delivery of device requires placement of vascular conduit or is through an established vascular graft
  • Receiving dialysis
  • History of hypercoagulability
  • Allergy to IV contrast
  • Serum creatinine level >2.0 mg/dL (or equivalent)
  • Patient is pregnant or nursing
  • Patient is currently enrolled in an investigational drug or device trial

Outcomes

Primary Outcomes

Safety- Number of Major Adverse Events

Time Frame: 30 day

The primary safety endpoint is defined as the incidence of Major Adverse Events (MAE) through 30 days after the index procedure.

Device Performance

Time Frame: 30 days

The primary performance endpoint is technical success defined as successful delivery and deployment of the device and freedom from open surgical conversion at 30 days.

Secondary Outcomes

  • All Cause Mortality(12 months)

Investigators

Sponsor
Endologix
Sponsor Class
Industry
Responsible Party
Sponsor

Study Sites (7)

Loading locations...

Similar Trials