S-ICD Electrode Registry

Not yet recruiting
Conditions
ICD Malfunction
Sudden Cardiac Death
Registration Number
NCT05713708
Lead Sponsor
Universitätsklinikum Köln
Brief Summary

The subcutaneous implantable cardioverter defibrillator (S-ICD) is established in the prevention of sudden cardiac death, but recently a defect of the electrode's Sense-B component was reported in some patients. This lead to inadequate shocks due to noise. The cause and incidence of this defect are unknown. The manufacturer issued a warning regarding this defect and a possible premature depletion of the battery in some device. Independent data on the incidence of these issues is limited. Further data is needed to better understand the risk of these complications.

Detailed Description

The S-ICD therapy is an established treatment for primary and secondary prevention of sudden cardiac death. Recently, a report was published about a previously described complication of the electrode, in which a defect of the so-called Sense-B component is reported in a few affected patients. The defect of this component of the electrode leads to electrode noise and subsequent inadequate shocks. The cause of the defect and the incidence are currently unknown. The manufacturer Boston Scientific has issued a warning and advisory.

In addition to this advisory regarding the electrode of the S-ICD, there is also an active warning from the manufacturer regarding the possible premature exhaustion of the S-ICD battery. This is caused by an error in the manufacturing process of approximately 40,000 devices produced in 2018. The manufacturer estimates the incidence of premature battery exhaustion to be 4-15% within 5 years. Independent data on the incidence of premature exhaustion of these devices were initially scarce. The investigators conducted a retrospective data query in 2021 to capture the incidence of battery exhaustion at that time. The incidence was 3.7% after 5 years. The average follow-up time was only 3 years. The incidence of premature exhaustion increased significantly after 3 years, so data with a longer follow-up period is urgently needed.

Given that these data already show that the incidence of premature exhaustion increases significantly after 3 years and the observation period in the retrospective data was on average slightly under 3 years, a further query of the existing clinical data on the battery as part of this register is also planned. This would allow for a more accurate estimate of the risk of premature battery depletion.

Recruitment & Eligibility

Status
NOT_YET_RECRUITING
Sex
All
Target Recruitment
1000
Inclusion Criteria
  • status post S-ICD insertion
Exclusion Criteria
  • no available follow-up data beyond 2 weeks after insertion

Study & Design

Study Type
OBSERVATIONAL
Study Design
Not specified
Primary Outcome Measures
NameTimeMethod
Sense B Electrode Failure3 months to 70 months after device insertion

Incidence of the Sense B Electrode Failure of the S-ICD Lead.

Secondary Outcome Measures
NameTimeMethod
Premature Battery Depletion3 months to 70 months after device insertion

Incidence of battery depletion of the S-ICD generator after insertion.

Trial Locations

Locations (1)

University Hospital Cologne

🇩🇪

Cologne, NRW, Germany