研究概览

简要总结

Objective: to evaluate the effectiveness, safety and level of satisfaction of parents and healthcare team with one innovative device for phototherapy to which a LED light mesh has been incorporated(a blanket), comparing it with the Ohmeda BiliBlanket Plus®. Methods: randomized controlled clinical trial with preterm or low birth weight infants requiring phototherapy. The study protocol was evaluated and accepted by the San Ignacio Hospital and "Pontificia Universidad Javeriana" research committee. Informed consent was requested from parents and the authors declared no conflict of interest. Sample size and allocation: using the STATA 12 program, a sample size of 64 patients was calculated(5% losses), 32 for each group, with significance level of 0.05, power of 80% and difference in bilirubin decreasing of at least 0,1 mg/dl/h with standard deviation of 0.14 mg/dl/h. A computer program randomized the allocation to the intervention and the concealment of the assignment was through sealed opaque envelopes. Seric bilirubin levels were taken at the start and at the end of phototherapy. Controls were done every 2 hours with the Bilicheck spectrophotometer device. Axillary temperature was measured at the beginning of phototherapy, at 30 min, 60 min, and every 2 hours during intervention. Physical examination and recording of side effects related were performed. Parents and health personnel answered a survey on comfort and perceptions with the use of the devices at the end of phototherapy. Analysis: Comparison of means was made for the decrease of bilirubin levels and body temperatures and chi-square for incidence of side effects and results of the survey.

详细描述

Background: Jaundice due to hyperbilirubinemia is the most common entity that requires medical management in neonates. About 50% of full-term newborns and 80% of premature infants develop jaundice. The effectiveness of phototherapy for neonatal jaundice depends on two factors: the spectrum of light emitted and the spectral irradiation of light. An ideal phototherapy device should have a broad light emission surface, in order to cover the maximum of the body surface in the horizontal plane, it should be durable, generate relatively little heat and provide a wavelength and light intensity (radiation) optimal (460-490nm and ≥ 30uw / cm2 / nm). The conventional phototherapy devices use fluorescent lamps that emit heat and require eye protectors, in addition to increasing water losses, which requires a permanent monitoring of vital functions and temperature. Its use imposes hospitalization and consequently mother-child separation, always traumatic. The most recent devices use light emitting diodes (LED light) as the light source. Objective: to evaluate the effectiveness, safety and level of satisfaction of parents and healthcare team with one innovative device for phototherapy to which a LED light mesh has been incorporated (one blanket), comparing it with the Ohmeda BiliBlanket Plus®, used at the Kangaroo Mother Care Program for intermittent ambulatory phototherapy. Methods: randomized controlled clinical trial with preterm or low birth weight infants that required phototherapy according to the guidelines of the American Academy of Pediatrics. The Protocol was evaluated and accepted by the "Hospital Universitario San Ignacio and Pontificia Universidad Javeriana research committee". Informed consent was requested from parents and the authors declared no conflict of interest. Sample size and allocation: using the STATA 12 program, a sample size of 64 patients was calculated, 32 for each group, with significance level of 0.05, power of 80% and difference in bilirubin decreasing of at least 0,1 mg/dl/h with standard deviation of 0.14 mg/dl/h and 5% of losses. A computer program randomized the allocation to the intervention and the concealment of the assignment was through sealed opaque envelopes. Seric bilirubin levels were taken at the start and at the end of phototherapy. Controls were done every 2 hours with bilicheck spectrophotometer device. Axillary temperature was measured at the beginning of phototherapy, at 30 min, 60 min, and every 2 hours during intervention. Physical examination, recording of side effects related, mortality and requirement of hospital treatment were performed. Parents and health personnel answered a survey on comfort and perceptions with the use of the devices at the end of phototherapy. Analysis: Comparison of means was made for the decrease of bilirubin levels and body temperatures and chi-square for incidence of side effects, hospitalization, mortality and results of the survey.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Parallel
主要目的
Treatment
盲法
None

入排标准

年龄范围
— 至 1 Month(Child)
性别
All
接受健康志愿者

入选标准

  • Low Risk Hyperbilirubinemia requiring phototherapy according to the guidelines of the American Academy of Pediatrics and recommendations for management of premature infants under 35 weeks, adopted by the Kangaroo Mother Care Ambulatory Program.
  • Start of jaundice at a chronological age greater than 48 hours.
  • Bilirubin level ≤18 mg / dL
  • Knowledge of blood groups of mother and child.
  • Parents are required to sign informed consent where they agree to participate in the study.

排除标准

  • Parents without availability to stay in the Kangaroo Mother Care Program for at least 6 hours.
  • Extensive skin lesions due to burns or sloughing at the physical examination prior to entering the study.

研究组 & 干预措施

Neomedlight Phototherapy Blanket

Experimental

Phototherapy with a fiber-optic device based on LED light administered intermittently for a total of 6 hours with periods of 2 hours in kangaroo position and pauses of 1 hour at the end of each period.

干预措施: Neomedlight Phototherapy Blanket (Device)

Ohmeda-Fiber Optic Phototherapy Blanket

Active Comparator

Phototherapy with a fiber-optic device: the Ohmeda fiber optic Phototherapy blanket

干预措施: Ohmeda-Fiber Optic Phototherapy Blanket (Device)

结局指标

主要结局

Bilirubin change rate in mg/dl/h

时间窗: through study completion, an average of 6 hours

Bilirubin lowering rate (initial seric bilirubin- final seric bilirubin/time of phototherapy).

次要结局

  • Patient temperature during phototherapy(through study completion, an average of 6 hours)
  • Incidence of skin lesions related to phototherapy(through study completion, an average of 6 hours)
  • Incidence of dehydration related to phototherapy(through study completion, an average of 6 hours)
  • Incidence of diarrhea related to phototherapy(through study completion, an average of 6 hours)
  • Incidence of other side effects related to phototherapy(through study completion, an average of 6 hours)
  • Perceptions of parents with the use of devices(At study completion, an average of 6 hours.)
  • Perceptions of the health team with the use of the devices(At study completion, an average of 6 hours.)
  • Hospitalization for hyperbilirubinemia(Between beginning of treatment and 72 hours after)
  • Mortality(Between beginning of treatment and 72 hours after)

研究者

发起方
Nathalie Charpak
申办方类型
Other
责任方
Sponsor Investigator
主要研究者

Nathalie Charpak

Director

Kangaroo Foundation

研究点 (2)

Loading locations...

相似试验