A Randomized Trial Comparing Two Forms of Immobilization of the Head for Fractioned Stereotactic Radiotherapy
试验速览
- 阶段
- 3 期
- 状态
- 已完成
- 入组人数
- 83
- 试验地点
- 2
- 主要终点
- Distance From Ideal to Center of GTC Frame and BrainLab Thermoplastic Mask With Respect to Average and Variability
研究概览
简要总结
Fractionated stereotactic radiotherapy (SRT) requires extremely precise and reproducible immobilization of the patient's head. This randomized study compares the efficacy of two commonly used forms of immobilization used for SRT.
详细描述
PRIMARY OBJECTIVE:
- To compare the efficacy of two commonly used, generally accepted forms of immobilization for the treatment of lesions in the head with fractionated stereotactic radiotherapy. Specifically, this study will use the ExacTRAC kilovoltage imaging system to assess positioning accuracy for these two systems.
OUTLINE:
Two routinely used methods of immobilization, which differ in their approach to reproduce head position from day-to-day are the Gill-Thomas-Cosman (GTC)frame and the BrainLab thermoplastic mask. The GTC frame fixates on the patient upper detention and thus us in direct mechanical contact with the cranium. The BrainLab mask is a two-part masking system custom-fitted to the front and back of the participant's head. After patients sign an IRB-approved informed consent form, eligible participants are either randomized to either the GTC frame or the mask for their course of SRT. Participants are treated as per standard procedure; however, prior to each treatment, a set of digital kV images (ExacTrac, BrainLabAB, Germany) is taken. These images are fused with reference digitally constructed radiographs obtained from treatment planning CT to yield lateral, longitudinal, and vertical deviations of isocenter and head rotations about respective axes. The primary endpoint of the study is to compare the two systems with respect to mean and standard deviations (SDs) using the distance to isocenter measure.
研究设计
- 研究类型
- Interventional
- 分配方式
- Randomized
- 干预模型
- Parallel
- 主要目的
- Treatment
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Patients who are eligible for fractionated stereotactic radiotherapy
排除标准
- •Severe claustrophobia
结局指标
主要结局
Distance From Ideal to Center of GTC Frame and BrainLab Thermoplastic Mask With Respect to Average and Variability
时间窗: Measurements taken during SRT
This study uses the ExacTRAC imaging system to assess positioning of frame or mask during SRT. Images yield lateral, longitudinal, and vertical deviations of the isocenter as well as head rotations about respective axes.
次要结局
未报告次要终点