Continuous Glucose Monitoring System Use in Pediatric Patients With Diabetes
试验速览
- 阶段
- 不适用
- 状态
- 招募中
- 发起方
- Sinocare
- 入组人数
- 78
- 试验地点
- 3
- 主要终点
- Accuracy
研究概览
简要总结
To verify and evaluate the efficacy and safety of i3 CGM system manufactured by Changsha Sinocare Inc. for continuous or regular monitoring of interstitial fluid glucose level in pediatric patients with diabetes.
详细描述
This clinical investigation is of a multi-center and self-controlled design.
This clinical investigation is planned to be conducted in 3 study sites, with a total of 78 subjects to be enrolled, of which no fewer than 15 cases are aged 2-5 years old, and no fewer than 60 cases are aged 6-17 years old. This clinical investigation is to verify the safety and efficacy of the investigational device. Referring to the number of pediatric patients with diabetes in the 3 study sites, it is expected that about 20 subjects will be enrolled in each site every month, and it will take about 2 months to close the site. This clinical investigation is expected to be completed in about 6 months from the first participant recruited.
The real-time glucose values measured by CGM system on the left and right abdomen are compared with the venous blood glucose value (EKF value) measured by EKF analyzer, respectively, and the consistency rate is analyzed.
研究设计
- 研究类型
- Observational
- 观察模型
- Other
- 时间视角
- Cross Sectional
入排标准
- 年龄范围
- 2 Years 至 17 Years(Child)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Aged 2 (inclusive) - 18 years (exclusive);
- •Clinically diagnosed with diabetes;
- •Being willing to wear the device continuously for 15 days and conduct blood collection for blood glucose testing in accordance with the requirements of this protocol;
- •Body weight not less than 10.0 kg;
- •Consent to participate in this clinical investigation, and sign the Informed Consent Form (ICF).
排除标准
- •Need to undergo magnetic resonance imaging (MRI) during the clinical investigation;
- •Diffuse subcutaneous nodules;
- •Acute complications of diabetes;
- •Abnormal coagulation function (coagulation cannot be too quick or too slow); the criteria for judging abnormalities are: The activated partial thromboplastin time (APTT) is higher than 1.5 times the upper limit of normal value or lower than the lower limit of normal value; or The prothrombin time (PT) is higher than 1.5 times the upper limit of normal value or lower than the lower limit of normal value
- •Participation in other clinical investigations within the past 1 month;
- •Where the investigator does not consider the patient suitable for enrollment.
结局指标
主要结局
Accuracy
时间窗: 0-15 days
percentage of i3 CGM values in the measuring range less than 4.4 mmol/L(80 mg/dL) within ± 1.1 mmol/L(20 mg/dL) of the EKF control.
次要结局
未报告次要终点