User Performance Evaluation of the INRatio® Prothrombin Time (PT) Monitoring System With a New INRatio Test Strip Designed for Low Sample Volume and Heparin Insensitivity
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 183
- 试验地点
- 4
- 主要终点
- Operate the INRatio Prothrombin Time (PT) Monitoring System utilizing the INRatio test strip newly designed for low sample volume and heparin insensitivity.
研究概览
简要总结
This multi-center study is designed to evaluate the ability of intended lay users (patients on oral anticoagulation therapy, OAT, or their caregivers) to 1) operate the INRatio Prothrombin Time (PT) Monitoring System utilizing the INRatio test strip newly designed for low sample volume and heparin insensitivity, and 2) obtain accurate results for the quantitative determination of International Normalized Ratio (INR) when self-testing using fingerstick capillary blood. Patients will be trained by their healthcare provider using the instructions for use and product labeling provided. The accuracy of the patient INR results will be assessed by comparison to the INR results obtained by the site's trained healthcare professional using the same INRatio system (from a separate fingerstick collected from the same patient at the point-of-care), and with the INR results obtained on venous plasma obtained from the same patient and analyzed at a central laboratory with the Sysmex CA-560 System INR reference method.
研究设计
- 研究类型
- Observational
- 观察模型
- Cohort
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Adults (18 years of age or older);
- •Ability to read and speak English;
- •Subjects on stable oral anticoagulation therapy with warfarin (at least 3 months);
- •Target INR interval between 2.0 and 4.5
- •Willing and able to provide written informed consent and comply with study procedures;
排除标准
- •Hematocrit less than 25 or greater than 55%
- •Lupus or antiphospholipid syndrome (APS)
- •Already participated in this study
- •A current INRatio PT Monitoring System User / Patient Self Tester
结局指标
主要结局
Operate the INRatio Prothrombin Time (PT) Monitoring System utilizing the INRatio test strip newly designed for low sample volume and heparin insensitivity.
时间窗: 8 weeks
次要结局
未报告次要终点