临床试验/NCT07321938
NCT07321938招募中不适用

Accuracy Evaluation of the Owlet OSS 3.0-M1 Sensor in Toddlers and Preschoolers

Owlet Baby Care, Inc.2 个研究点 分布在 1 个国家目标入组 90 人开始时间: 2025年12月15日最近更新:
干预措施

试验速览

阶段
不适用
状态
招募中
入组人数
90
试验地点
2
主要终点
Pulse rate accuracy

研究概览

简要总结

The purpose of the study is to collect information from the Owlet OSS 3.0-M1 Sensor in toddlers and preschoolers to evaluate accuracy.

详细描述

The objective of this study is to collect evidence to verify the accuracy of the Owlet OSS 3.0-M1 Sensor to measure SpO2 and pulse rate in children 19-60 months in a simulated use environment as compared to an FDA cleared reference pulse oximeter.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
19 Months 至 60 Months(Child)
性别
All
接受健康志愿者

入选标准

  • Subject is between 19 months and ≤ 60 months of age.
  • Subject's current body weight is ≥ 6 lbs.
  • Subject/legal guardian is fluent in both written and spoken English.
  • Subject is willing and able to follow study requirements
  • Legal guardian has provided consent.

排除标准

  • Subject has preexisting cardiovascular or respiratory disease or conditions.
  • Subjects with abnormalities (e.g., limb deformities, open wounds, severe edema) at the planned application sites that would interfere with system measurements.
  • The device(s) are unable to fit properly without risk of injury (e.g., hook of VELCRO is against the skin, wrap is too tight and constricting).
  • Subject has skin irritation at the location of device placement.
  • Subject has known allergies to adhesive.

研究组 & 干预措施

Owlet OSS 3.0-M1 Sensor

Experimental

Owlet OSS 3.0-M1 Sensor accuracy evaluation compared to a reference pulse oximeter.

干预措施: Application and monitoring with an Owlet OSS 3.0-M1 Sensor (Device)

Owlet OSS 3.0-M1 Sensor

Experimental

Owlet OSS 3.0-M1 Sensor accuracy evaluation compared to a reference pulse oximeter.

干预措施: Reference Pulse Oximeter (Device)

结局指标

主要结局

Pulse rate accuracy

时间窗: 2 hours

Average root mean square (ARMS) error between the pulse rate measured by the Owlet OSS 3.0-M1 vs. a reference pulse oximeter

SpO2 Accuracy

时间窗: 2 hours

The average root mean square difference between the Owlet OSS 3.0-M1 Sensor vs. readints from an FDA cleared reference pulse oximeter.