Volar Plates for Distal Radius : Can Epiphyseal Screw Length be Predicted

University Hospital, Brest1 个研究点 分布在 1 个国家目标入组 41 人开始时间: 2018年1月25日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
41
试验地点
1
主要终点
Main objective

研究概览

简要总结

Background Fractures of the distal radius are among the most common. Major complications, including irritation and rupture of the extensor tendons, may occur if epiphyseal screws of inappropriate length are used.

Questions/Purposes The main objective of the present study was to determine whether the optimal epiphyseal screw length can be determined by reference to the diaphyseal screw length.

Methods Forty CT scans were semi-automatically segmented. A 3D model of the volar plate was affixed to each distal radius with simulation. The maximum lengths of the diaphyseal screws, as well as the four distal epiphyseal screws, were measured. Linear regression analysis was performed.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Retrospective

入排标准

年龄范围
17 Years 至 85 Years(Child, Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • All CT-scans available in our database

排除标准

  • Any pathology that could affect the radius morphology
  • Refuse to participate

结局指标

主要结局

Main objective

时间窗: The measures were performed after the CT-Scans were realized. The CT-Scans were performed from October 2016 to February 2018. The measures on the CT-Scans were realized in june 2018

The main objective of this study was twofold: 1) to assess the anatomical correlation between diaphysis thickness in the diaphyseal screw region of volar plates and epiphyseal thickness in lateral plane with linear regression analysis for each epiphyseal screw length with regard to the diaphysis screw length, and 2) to evaluate whether the optimal epiphyseal screw lengths can be predicted by reference to the diaphyseal screw length when diaphyseal screws are first positioned. From the CT-scans, Reconstructions in 3D of the radius were realized. Those reconstructions were used in a computer-simulation study. A Volar plate was affixed to the radius using a specific software. Maximum length for each screw located either in epiphysis or diaphysis was measured in mm. This length corresponded with the distance between the entry point of the screw and the opposite point belonging to the wrist along the line perpendicular to the plate.