Optimal Number of To-and-fro Motion in EUS-guided Fine Needle Aspiration for Pancreatic Masses in Terms of Suction or Without Suction: Prospective Randomized Single Blinded Trial
试验速览
- 阶段
- 3 期
- 状态
- 已完成
- 入组人数
- 193
- 试验地点
- 2
- 主要终点
- Diagnostic accuracy
研究概览
简要总结
EUS-FNA is the standard of care for diagnosing pancreatic masses. According to the operator's preference, EUS-FNA with suction or without suction technique has been used. To date, little is known about optimal number of to-and-fro motion during each needle pass. As usual 10-15 to-and-fro motions was used for EUS-FNA with suction technique. Theoretically, more number of to-and-fro motions may be required in EUS-FNA without suction. In this circumstance, the contamination of blood in specimen and possible adverse event may occur. To determine optimal number of to-and-fro motion in EUS-guided FNA for pancreatic masses in terms of with suction or without suction, this prospective single-blinded randomized trial was conducted.
详细描述
For this prospective single-blinded randomized trial, two 10cc syringes was prepared. Two syringe was provided as same performance. However, one syringe is falsely made as suction syringe (fake one). During EUS-FNA, two syringe is randomly assigned and loaded by a nurse. Thus, operator was blinded which syringe is with suction or without suction.
First hypothesis: There is difference in the diagnostic accuracy of pancreatic masses according to EUS-FNA with suction or without suction/ different number of to-and-fro motion.
Second hypothesis: There is difference in the amount of blood in samples obtained by EUS-FNA with suction or without suction/ different number of to-and-fro motion.
研究设计
- 研究类型
- Interventional
- 分配方式
- Randomized
- 干预模型
- Parallel
- 主要目的
- Diagnostic
- 盲法
- Double (Investigator, Outcomes Assessor)
入排标准
- 年龄范围
- 18 Years 至 85 Years(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Age > 19 years Solid pancreatic mass lesions
排除标准
- •Age < 19 years Coagulopathy Unable to consent
结局指标
主要结局
Diagnostic accuracy
时间窗: 1 year
The diagnostic accuracy of pancreatic masses at each sample according to EUS-FNA with suction or without suction/ different number of to-and-fro motion will be measured. Diagnostic accuracy of EUS-FNA with suction and 10, 15, and 20 to-and-fro motion vs.Diagnostic accuracy of EUS-FNA without suction and 10, 15, and 20 to-and-fro motion.
次要结局
- Blood contamination in each sample(1 year)
研究者
Do Hyun Park
Assistant Professor
Asan Medical Center