临床试验/NCT01726114
NCT01726114已完成不适用

A Clinical Study to Evaluate the Performance of the C8 MediSensors Optical Glucose Monitor™ During On-Site Testing and Real Time Use Monitoring

C8 MediSensors, Inc.2 个研究点 分布在 1 个国家目标入组 86 人开始时间: 2012年6月1日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
86
试验地点
2
主要终点
Performance of the test device in comparison to the reference device

研究概览

简要总结

The purpose of this study is to compare glucose measurements from the C8 MediSensors non-invasive Optical Glucose Monitor™ System to blood glucose measurements from the Yellow Springs Instrument(YSI) blood glucose analyzer.

详细描述

This is a prospective, non-randomized, open label study to evaluate the performance of the C8 MediSensors non-invasive Optical Glucose Monitor™ on individuals with diabetes mellitus. Capillary blood samples will be obtained throughout a daily study session; the study session may last for up to 12 hours. Blood glucose will be measured on a YSI blood glucose analyzer. Subjects may adjust food intake and insulin administration so device performance is evaluated across a broad range of glucose concentrations. Glucose measurements will be time matched and paired for comparison. Skin effects and adverse events will be evaluated.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • 18 years of age or older.
  • Understands and agrees to comply with study instructions.
  • Read, understood, signed and dated the Informed Consent Form.

排除标准

  • Extensive skin changes or diseases that preclude wearing the test device at the proposed wear site(s) (e.g., extensive psoriasis, recent burns or severe sunburn, extensive eczema, extensive scarring, extensive tattoos, dermatitis herpetiformis).
  • Employed by a company within the diabetes field other than the study sponsor.
  • Medical history or any condition that may, in the opinion of the investigator, compromise the subject's ability to participate.

结局指标

主要结局

Performance of the test device in comparison to the reference device

时间窗: 6 months

The primary objective is to characterize the accuracy of the non-invasive glucose measurements from the Optical Glucose Monitor™ to blood glucose measurements on a sample size of approximately 30 subjects with 1000 time matched, paired data points.

次要结局

未报告次要终点

研究者

发起方
C8 MediSensors, Inc.
申办方类型
Industry
责任方
Sponsor