A Prospective Study of Airless Tubing in an Inpatient Acute Hemodialysis Unit in Hospitalized Patients in a Large Medical Center
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 338
- 试验地点
- 1
- 主要终点
- Clotting of Extracorporeal Dialysis Circuit
研究概览
简要总结
The introduction of unfractionated heparin (UFH), which prevents clotting of the extracorporal circuit, was one of the key advances that led to the rapid development and expansion of hemodialysis services. However, anticoagulation during hemodialysis of the patient at high risk for bleeding remains a frequently encountered problem in both inpatient and outpatient dialysis practice.
Streamline bloodlines are designed to eliminate blood-air contact. This is thought to help reduce heparin use and decrease clotting rates. The goal of this study was to prospectively examine impact of the Streamline airless blood tubing set, in an inpatient setting, on dialysis circuit clotting rates, anticoagulation use, and dialysis efficiency.
详细描述
The introduction of unfractionated heparin (UFH), which prevents clotting of the extracorporeal circuit, was one of the key advances that led to the rapid development and expansion of hemodialysis use, and remains the mainstay in hemodialysis practice today.
However, anticoagulation during hemodialysis of the patient at high risk for bleeding remains a frequently encountered problem in the nephrology practice. The need for anticoagulation to prevent clotting of the extracorporeal blood circuit and the need to prevent anticoagulation related bleeding complications in the patient has led to the development of numerous strategies; the safest from a bleeding standpoint being anticoagulant-free hemodialysis.
Streamline® bloodlines (Medisystems® Corporation, Lawrence, MA) are designed to eliminate blood-air contact. A pressure pod measures arterial and venous pressures without any blood-air contact. The venous chamber is run without an air gap. It is also designed so that blood flows in a circular vortex manner. This airless system is thought to provide several benefits: improved dialysis efficiency and blood flow rates, reduced heparin use and clotting rates.
The goal of this study was to prospectively examine the Streamline® airless tubing system in an inpatient setting and its association with clotting rates, and dialysis efficiency.
研究设计
- 研究类型
- Observational
- 观察模型
- Cohort
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •All adult non-pregnant patients requiring inpatient hemodialysis at our center during the recruitment period until the target sample of dialysis sessions (n=1200) was reached.
排除标准
- •Pregnant females
结局指标
主要结局
Clotting of Extracorporeal Dialysis Circuit
时间窗: During the dialysis session, approximately 211 minutes
Each dialysis session was prospectively observed for Clotting of Extracorporeal Dialysis Circuit. Events were defined as interruption of hemodialysis session, loss of the hemodialysis circuit, or inability to return blood to the patient upon rinse back.