A Prospective Study of Airless Tubing in an Inpatient Acute Hemodialysis Unit in Hospitalized Patients
- Conditions
- End Stage Renal Failure on DialysisComplication of DialysisBlood Coagulation Disorders
- Registration Number
- NCT02086682
- Lead Sponsor
- Marie Hogan
- Brief Summary
The introduction of unfractionated heparin (UFH), which prevents clotting of the extracorporal circuit, was one of the key advances that led to the rapid development and expansion of hemodialysis services. However, anticoagulation during hemodialysis of the patient at high risk for bleeding remains a frequently encountered problem in both inpatient and outpatient dialysis practice.
Streamline bloodlines are designed to eliminate blood-air contact. This is thought to help reduce heparin use and decrease clotting rates. The goal of this study was to prospectively examine impact of the Streamline airless blood tubing set, in an inpatient setting, on dialysis circuit clotting rates, anticoagulation use, and dialysis efficiency.
- Detailed Description
The introduction of unfractionated heparin (UFH), which prevents clotting of the extracorporeal circuit, was one of the key advances that led to the rapid development and expansion of hemodialysis use, and remains the mainstay in hemodialysis practice today.
However, anticoagulation during hemodialysis of the patient at high risk for bleeding remains a frequently encountered problem in the nephrology practice. The need for anticoagulation to prevent clotting of the extracorporeal blood circuit and the need to prevent anticoagulation related bleeding complications in the patient has led to the development of numerous strategies; the safest from a bleeding standpoint being anticoagulant-free hemodialysis.
Streamline® bloodlines (Medisystems® Corporation, Lawrence, MA) are designed to eliminate blood-air contact. A pressure pod measures arterial and venous pressures without any blood-air contact. The venous chamber is run without an air gap. It is also designed so that blood flows in a circular vortex manner. This airless system is thought to provide several benefits: improved dialysis efficiency and blood flow rates, reduced heparin use and clotting rates.
The goal of this study was to prospectively examine the Streamline® airless tubing system in an inpatient setting and its association with clotting rates, and dialysis efficiency.
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 338
- All adult non-pregnant patients requiring inpatient hemodialysis at our center during the recruitment period until the target sample of dialysis sessions (n=1200) was reached.
- Pregnant females
- Children
Study & Design
- Study Type
- OBSERVATIONAL
- Study Design
- Not specified
- Primary Outcome Measures
Name Time Method Clotting of Extracorporeal Dialysis Circuit During the dialysis session, approximately 211 minutes Each dialysis session was prospectively observed for Clotting of Extracorporeal Dialysis Circuit. Events were defined as interruption of hemodialysis session, loss of the hemodialysis circuit, or inability to return blood to the patient upon rinse back.
- Secondary Outcome Measures
Name Time Method