NOWDx Test for the Diagnosis of Syphilis

NOWDiagnostics, Inc.6 个研究点 分布在 1 个国家目标入组 1,535 人开始时间: 2021年10月4日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
1,535
试验地点
6
主要终点
Compare serological status to comparator method

研究概览

简要总结

This study is designed to compare the performance of the NOWDx Syphilis Test to a currently marketed device. The intent is to show the rapid test device is comparable to the currently marketed device. The NOWDx Syphilis Test is intended for qualitatively detecting the presence or absence of human antibodies to syphilis in human whole blood to aid in the diagnosis of infection caused by Treponema pallidum.

详细描述

The objective of this study is to establish the performance characteristics of the NOWDx Syphilis Test based on comparison to the BioRad Bioplex 2200 series Syphilis Total (IgG/IgM), Wampole Impact RPR Test Kit, and to the Serodia Treponema pallidum particle agglutination assay (TP-PA).

The external clinical study along with in-house analytical studies will demonstrate the efficacy of the NOWDx Syphilis Test as an aid in the diagnosis of syphilis infection for at home testing and point of care testing sites. Participants will self test and be tested by Clinical Laboratory Improvement Amendments (CLIA) Waived operators with the NOWDx Syphilis Test at independent sites in WA, AZ, and NV. Two sample types will be tested with the NOWDx Syphilis Test for each participant: fingerstick whole blood and venous whole blood. Sera from each participant will be tested at an independent reference laboratory.

The NOWDx Syphilis Test will be evaluated in diverse populations of sexually active persons who self-select for syphilis testing and expectant mothers.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 64 Years(Adult)
性别
All
接受健康志愿者

入选标准

  • 未提供

排除标准

  • 未提供

结局指标

主要结局

Compare serological status to comparator method

时间窗: through study completion, an average of 6 months

The comparator method used in this study is an algorithm comprised of results from three different commercially available syphilis assays: an FDA-cleared treponemal immunoassay, an FDA-cleared non-treponemal assay (RPR assay) and a second treponemal assay (TP-PA).

Evaluate test reproducibility by lay users

时间窗: through study completion, an average of 6 months

Test lay user reproducibility using prepared samples for % agreement with expected results.

次要结局

未报告次要终点