Annuloplasty Rings and Band Post-Market Clinical Follow-Up (ARB-PMCF) Study
试验速览
- 阶段
- 不适用
- 状态
- 招募中
- 入组人数
- 550
- 试验地点
- 33
- 主要终点
- Percentage of Participants in Each Mitral Treatment Group with Freedom from Reoperation or Transcatheter Reintervention for Mitral Regurgitation through 5 Years Post-Implant
研究概览
简要总结
ARB-PMCF is a multicenter, observational study of the safety and performance of Abbott annuloplasty devices used in surgical repair of mitral and tricuspid valve regurgitation. The devices included in this study are the SJM™ Rigid Saddle Ring and SJM Séguin Annuloplasty Ring, indicated for mitral valve repair, and the SJM Tailor™ Annuloplasty Ring and SJM Tailor Annuloplasty Band, indicated for mitral or tricuspid repair. Participants will be enrolled prior to undergoing mitral or tricuspid valve repair surgery including an Abbott annuloplasty implant and will complete annual follow-up visits through five years from implant. The study is being conducted to meet post-market clinical follow-up requirements of the European Union Medical Device Directives.
详细描述
The ARB-PMCF Study has a prospective, multicenter, observational, parallel group design in which up to 550 subjects will be followed through five years from implant of an Abbott annuloplasty device, with or without concomitant procedures, in five treatment groups, including primary mitral disease repair (N=200) with the Rigid Saddle Ring, Séguin Ring or full Tailor Ring without cut zone removal; secondary mitral disease repair (N=200) with the Rigid Saddle Ring, Séguin Ring or full Tailor Ring; primary tricuspid disease repair with the full Tailor Ring (N=50); secondary tricuspid disease repair with the full Tailor Ring (N=50); and primary tricuspid disease repair with the Tailor Band or partial Tailor Ring with cut zone removed (N=50). Study enrollment will occur at up to 25 global centers with approximately half or more of all implanted subjects being from European sites. The study will enroll adult subjects expected to be implanted with an Abbott annuloplasty device within 90 days who meet all other eligibility requirements. Participants will complete annual in-clinic or telephone follow-up visits through five years from their annuloplasty implant. Data collected will include adverse events, cardiac medication usage, New York Heart Association functional classification (a measure of the severity of heart failure symptoms) and echocardiographic assessments of regurgitation through the repaired valve.
研究设计
- 研究类型
- Observational
- 观察模型
- Cohort
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Subject is expected to undergo cardiac surgery in ≤90 days including at least one of the following:
- •implant of a Rigid Saddle Ring, Séguin Ring or a full Tailor Ring without cut zone removal for MR repair
- •implant of a full Tailor Ring without cut zone removal for TR repair, or
- •implant of a Tailor Band or partial Tailor Ring with cut zone removed for primary TR repair.
- •Subject's cardiac surgery will be performed by a study investigator.
- •Subject will be ≥18 years old at the time of their annuloplasty implant(s).
- •Subject provides written informed consent and agrees to comply with all required study visits and procedures.
排除标准
- •Subject is below the age of legal consent in the applicable jurisdiction or otherwise lacks legal authority to provide informed consent.
- •Subject is unable to read or write or has a mental illness or disability that impairs their ability to provide written informed consent.
- •Subject is expected to have active endocarditis at the time of their Abbott annuloplasty device implant(s).
- •Subject cannot be unambiguously assigned to a treatment group due to missing data on repair indication(s) or the model and configuration (cut zone removed or not) of their Abbott annuloplasty implant(s).
- •Subject is participating in another clinical investigation including any treatment outside the investigative site's usual standard of care.
- •Subject has anomalous anatomy or medical, surgical, psychiatric or social history or conditions that, in the investigator's opinion, would limit the subject's ability to participate in the clinical investigation or to comply with follow-up requirements
研究组 & 干预措施
Group 1: Primary mitral disease repair surgery with an Abbott annuloplasty ring implant
Group 1 will enroll 200 subjects undergoing surgical repair of primary mitral regurgitation that includes annuloplasty with an Abbott SJM Rigid Saddle Ring, Séguin Ring or full Tailor Ring without cut zone removal. Enrollment must include at least 50 subjects implanted with each ring model. In primary mitral regurgitation, backflow through the closed valve is caused by disease intrinsic to the mitral valve tissue itself.
干预措施: Abbott SJM Rigid Saddle Ring (Device)
Group 1: Primary mitral disease repair surgery with an Abbott annuloplasty ring implant
Group 1 will enroll 200 subjects undergoing surgical repair of primary mitral regurgitation that includes annuloplasty with an Abbott SJM Rigid Saddle Ring, Séguin Ring or full Tailor Ring without cut zone removal. Enrollment must include at least 50 subjects implanted with each ring model. In primary mitral regurgitation, backflow through the closed valve is caused by disease intrinsic to the mitral valve tissue itself.
干预措施: Abbott SJM Séguin Ring (Device)
Group 1: Primary mitral disease repair surgery with an Abbott annuloplasty ring implant
Group 1 will enroll 200 subjects undergoing surgical repair of primary mitral regurgitation that includes annuloplasty with an Abbott SJM Rigid Saddle Ring, Séguin Ring or full Tailor Ring without cut zone removal. Enrollment must include at least 50 subjects implanted with each ring model. In primary mitral regurgitation, backflow through the closed valve is caused by disease intrinsic to the mitral valve tissue itself.
干预措施: Abbott SJM Tailor Ring (Device)
Group 2: Secondary mitral disease repair surgery with an Abbott annuloplasty ring implant
Group 2 will enroll 200 subjects undergoing surgical repair of secondary mitral regurgitation that includes annuloplasty with an Abbott SJM Rigid Saddle Ring, Séguin Ring or full Tailor Ring without cut zone removal. Enrollment must include at least 50 subjects implanted with each ring model. In secondary mitral regurgitation, backflow through the closed valve is secondary to diseases of the surrounding myocardium, rather than caused by disease of the valve tissue itself.
干预措施: Abbott SJM Rigid Saddle Ring (Device)
Group 2: Secondary mitral disease repair surgery with an Abbott annuloplasty ring implant
Group 2 will enroll 200 subjects undergoing surgical repair of secondary mitral regurgitation that includes annuloplasty with an Abbott SJM Rigid Saddle Ring, Séguin Ring or full Tailor Ring without cut zone removal. Enrollment must include at least 50 subjects implanted with each ring model. In secondary mitral regurgitation, backflow through the closed valve is secondary to diseases of the surrounding myocardium, rather than caused by disease of the valve tissue itself.
干预措施: Abbott SJM Séguin Ring (Device)
Group 2: Secondary mitral disease repair surgery with an Abbott annuloplasty ring implant
Group 2 will enroll 200 subjects undergoing surgical repair of secondary mitral regurgitation that includes annuloplasty with an Abbott SJM Rigid Saddle Ring, Séguin Ring or full Tailor Ring without cut zone removal. Enrollment must include at least 50 subjects implanted with each ring model. In secondary mitral regurgitation, backflow through the closed valve is secondary to diseases of the surrounding myocardium, rather than caused by disease of the valve tissue itself.
干预措施: Abbott SJM Tailor Ring (Device)
Group 3: Primary tricuspid disease repair surgery with a full Tailor Ring implant
Group 3 will enroll up to 50 subjects undergoing surgical repair of primary tricuspid regurgitation that includes annuloplasty with a full Abbott SJM Tailor Ring without cut zone removal. In primary tricuspid regurgitation, backflow through the closed valve is caused by disease intrinsic to the tricuspid valve tissue itself.
干预措施: Abbott SJM Tailor Ring (Device)
Group 4: Secondary tricuspid disease repair surgery with a full Tailor Ring implant
Group 4 will enroll up to 50 subjects undergoing surgical repair of secondary tricuspid regurgitation that includes annuloplasty with a full Abbott SJM Tailor Ring without cut zone removal. In secondary tricuspid regurgitation, backflow through the closed valve is secondary to diseases of the surrounding myocardium, rather than caused by disease of the valve tissue itself.
干预措施: Abbott SJM Tailor Ring (Device)
Group 5: Primary tricuspid disease repair surgery with a partial Tailor Ring or Tailor Band implant
Group 4 will enroll up to 50 subjects undergoing surgical repair of secondary tricuspid regurgitation that includes posterior annuloplasty with either a partial Abbott SJM Tailor Ring with cut zone removed or an Abbott SJM Tailor Band. In primary tricuspid regurgitation, backflow through the closed valve is caused by disease intrinsic to the tricuspid valve tissue itself.
干预措施: Abbott SJM Tailor Ring (Device)
Group 5: Primary tricuspid disease repair surgery with a partial Tailor Ring or Tailor Band implant
Group 4 will enroll up to 50 subjects undergoing surgical repair of secondary tricuspid regurgitation that includes posterior annuloplasty with either a partial Abbott SJM Tailor Ring with cut zone removed or an Abbott SJM Tailor Band. In primary tricuspid regurgitation, backflow through the closed valve is caused by disease intrinsic to the tricuspid valve tissue itself.
干预措施: Abbott SJM Tailor Band (Device)
结局指标
主要结局
Percentage of Participants in Each Mitral Treatment Group with Freedom from Reoperation or Transcatheter Reintervention for Mitral Regurgitation through 5 Years Post-Implant
时间窗: 5 years
Percentage of participants in each mitral treatment group (Groups 1 and 2) who are free from reoperation or transcatheter intervention for mitral regurgitation through 5 years after their Abbott annuloplasty implant, estimated from time-to-event data using the Kaplan-Meier estimator. This outcome will be reported separately for Group 1 and Group 2.
Percentage of Participants in Each Treatment Group with Freedom from All-Cause Mortality through 5 Years Post-Implant
时间窗: 5 years
Percentage of participants in each study treatment group (Groups 1 through 5) who are free of death from any cause through 5 years after their Abbott annuloplasty device implant, estimated from time-to-event data using the Kaplan-Meier estimator. This outcome will be reported for each treatment group separately.
Percentage of Participants in Each Tricuspid Treatment Group with at least a 1-Class Reduction in New York Heart Association (NYHA) Functional Classification at 1 Year Post-Implant.
时间窗: 1 year
This outcome will be reported for each tricuspid repair group (Groups 3 through 5) separately. NYHA functional classification is a widely used indicator of the severity of heart failure symptoms with four classes: Class I. No limitation of physical activity. Ordinary physical activity does not cause undue fatigue, palpitation, or dyspnea (shortness of breath). Class II. Slight limitation of physical activity. Comfortable at rest. Ordinary physical activity results in fatigue, palpitation, or dyspnea. Class III. Marked limitation of physical activity. Comfortable at rest. Less than ordinary activity causes fatigue, palpitation, or dyspnea. Class IV. Unable to carry on any physical activity without discomfort. Symptoms of heart failure at rest. If any physical activity is undertaken, discomfort increases.
次要结局
- Number of Participants in Each Treatment Group by Grade of Regurgitation in the Valve Undergoing Repair, Pre-Operatively, at Discharge, and at 1, 3 and 5 Years Post-Implant(Pre-Operative Baseline, within 60 days post-implant, and 1, 3 and 5 years post-implant)
- Number of Participants in Each Treatment Group by NYHA Functional Class, Pre-Operatively and at Discharge, 1, 3 and 5 Years Post-Implant.(Pre-Operative Baseline, within 60 days post-implant, and 1, 3 and 5 years post-implant)