Evaluation of the Impact of SpO2 Averaging Time on Performance of an Automatic FiO2 Control System: a Randomized Study
- Conditions
- Very Low Birth Weight InfantNeonatal Respiratory Distress
- Interventions
- Device: Fabian ventilator with PRICO system (Acutronic Medical Systems AG, Hirzel, Switzerland); SpO2 averaging time set to 16 sDevice: Fabian ventilator with PRICO system (Acutronic Medical Systems AG, Hirzel, Switzerland); SpO2 averaging time set to 4to6 sDevice: Fabian ventilator with PRICO system (Acutronic Medical Systems AG, Hirzel, Switzerland); SpO2 averaging time set to 10 s
- Registration Number
- NCT05274386
- Lead Sponsor
- Czech Technical University in Prague
- Brief Summary
The aim of the study is to determine the preferred oximeter averaging setting during automated control of FiO2 (A-FiO2) in infants receiving respiratory support and supplemental oxygen.
- Detailed Description
There are 7 different averaging time settings available with PRICO (Acutronic Medical Systems AG, Hirzel, Switzerland), but after over a year of experience there is no clear clinical impression of the best setting. Therefore, a small systematic study is needed to determine the optimal guidelines.
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 10
- All VLBW on respiratory support and oxygen requirements after 2 weeks of age in the NICU are eligible after parental informed consent is obtained.
- Parental informed consent is not obtained
- Recording device for automated control of FiO2 is not available
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- CROSSOVER
- Arm && Interventions
Group Intervention Description Averaging time 16 s Fabian ventilator with PRICO system (Acutronic Medical Systems AG, Hirzel, Switzerland); SpO2 averaging time set to 16 s The SpO2 averaging time will be set to 16 s for the next 12 hours. All elements of care will be as routinely used, except the changing of the averaging time setting. Averaging time 4to6 s Fabian ventilator with PRICO system (Acutronic Medical Systems AG, Hirzel, Switzerland); SpO2 averaging time set to 4to6 s The SpO2 averaging time will be set to 4to6 s for the next 12 hours. All elements of care will be as routinely used, except the changing of the averaging time setting. Averaging time 10 s Fabian ventilator with PRICO system (Acutronic Medical Systems AG, Hirzel, Switzerland); SpO2 averaging time set to 10 s The SpO2 averaging time will be set to 10 s for the next 12 hours. All elements of care will be as routinely used, except the changing of the averaging time setting.
- Primary Outcome Measures
Name Time Method Percent time in SpO2 target range 20 days of intervention on average Percent time in the intended SpO2 target range (compliance), at SpO2 \<86% (safety), and at SpO2 \>98% (safety).
Period with SpO2 higher than the target range with FiO2 =0.21 will be included in the target range compliance and excluded from time above the target range.
- Secondary Outcome Measures
Name Time Method Effect of set SpO2 target range and mode of ventilation on the percent time in SpO2 target range, at SpO2 <86%, and at SpO2 >98%. 20 days of intervention on average A general linear model will be used controlling for target range (based on gestational age), mode of ventilation (e.g. High-frequency oscillatory ventilation, Continuous mandatory ventilation, Continuous positive airway pressure, High-flow nasal cannula, and Nasal intermittent positive pressure ventilation), and subject.
Trial Locations
- Locations (1)
Motol University Hospital
🇨🇿Prague, Czechia