Preliminary Evaluation of the Artificial Intelligence Software VX1+ for the Detection and Automatic Tagging on 3D Mapping of Spatio-temporal Dispersion Areas During Atrial Fibrillation or Atrial Tachycardia Ablation Procedures.
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 40
- 试验地点
- 2
- 主要终点
- Perioperative reliability of VX1+ in the real-time detection of electrograms in AF or AT exhibiting a spatio-temporal dispersion, and auto-tagging of these dispersion areas on 3D electroanatomical maps.
研究概览
简要总结
Interventional, prospective, non-randomized, single-center, non-controlled clinical investigation as part of the premarket clinical evaluation of VX1+ medical device, aiming to verify that VX1+ in bidirectional configuration with auto-tagging function is both ergonomic and reliable for dispersed electrograms detection and automatic tagging on 3D-map in real-time.
详细描述
Ablation of Atrial Fibrillation (AF) and Atrial Tachycardia (AT) is typically performed in predetermined anatomic regions of the left atrium (pulmonary veins isolation) and may be supplemented by a tailored ablation approach in order to identify areas of interest specific to the patient. The VX1+ medical device is intended to assist operators in the manual or automatic real-time annotation of 3D electroanatomical maps of the heart for the presence of multipolar intra-cardiac atrial electrograms exhibiting spatio-temporal dispersion during ablation procedures.
This study will specifically allow to evaluate the peroperative performance of VX1+ automatic tagging function in bi-directional configuration with a compatible 3D mapping system.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Diagnostic
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Patient aged 18 years or older.
- •Patient candidate for catheter ablation of atrial tachycardia, or paroxysmal or persistent atrial fibrillation, de novo or after one or several previous ablation procedures.
- •Continuous anticoagulation for more than 4 weeks before ablation.
- •Patient giving his signed consent form to participate in the clinical study.
- •Patient affiliated to the French social security
排除标准
- •Contraindication to AF/AT catheter ablation.
- •Major bleeding disorder.
- •Contraindication to anticoagulation (Heparin, Warfarin or novel oral anticoagulants (NOAC)) or lack of anticoagulation for 4 weeks prior to the procedure.
- •Presence of a left atrium (LA) thrombus on transesophageal echocardiography (TEE) prior to the procedure.
- •Patient who is or could potentially be pregnant.
- •Person deprived of liberty or under guardianship.
- •Patient's refusal to participate in the study.
结局指标
主要结局
Perioperative reliability of VX1+ in the real-time detection of electrograms in AF or AT exhibiting a spatio-temporal dispersion, and auto-tagging of these dispersion areas on 3D electroanatomical maps.
时间窗: During procedure
Comparison between the automatically annotated areas with VX1+ and those annotated manually following the operator's visual analysis (true positive, false positive, false negative).
次要结局
- Accuracy of VX1+ in estimating cardiac cycle lengths(During procedure)
- Preliminary safety data on VX1+ perioperatively in terms of device deficiencies, adverse events and mapping time(From the date of hospitalization for the ablation procedure until the date of discharge, up to 1 week)
- Operators' satisfaction regarding the ergonomics of VX1+(Through study completion, assessed up to 4 months)