Comparison of Easy urodynamic study and standard urodynamic study in measuring intravesical pressure
Phase 3
Completed
- Conditions
- neurogenic bladderupper urinary tract deteriorateNeurogenic bladderUpper urinary tract deteriorateUrodynamicManometry
- Registration Number
- TCTR20220515004
- Brief Summary
The results suggest that the easy UD could be used to measure bladder pressure (Pves) similarly to standard UD and E Pves cut-off more than 34 cmH2O can help identify patients with risk for upper tract deterioration (Pdet more than 30 cmH2O) who may benefit from early standard UD re-evaluation and closer follow-up
- Detailed Description
Not available
Recruitment & Eligibility
- Status
- Completed
- Sex
- All
- Target Recruitment
- 30
Inclusion Criteria
patients with history of neurogenic bladder who had indications for UDS
Exclusion Criteria
urinary tract infection
incontinence at the time of performing UDS or history of urethral stricture
Study & Design
- Study Type
- Interventional
- Study Design
- Not specified
- Primary Outcome Measures
Name Time Method intravesicular pressure by easy UD (E Pves) while doing easy UDS cmH2O
- Secondary Outcome Measures
Name Time Method relationship between E Pves, Pves and P abd while doing Easy UDS and standard UDS cmH2O