Feasibility and Safety of a Closed Loop Insulin Delivery System (Aka AAGC) With an Artificially Induced Calibration Error During the Overnight Period
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 9
- 试验地点
- 2
- 主要终点
- Amount of time spent in the target range defined as 3.8-8.8 mmol/l (70 to 160 mg/dL) YSI glucose during in-clinic overnight visits
研究概览
简要总结
Currently, patients diagnosed with type 1 diabetes rely on either finger stick or sensor glucose readings when making their insulin dosage decisions. Designing a computerized system that mimics the way insulin is produced naturally in a person who does not have type 1 diabetes holds many challenges; all of which cannot be addressed in just one study. The purpose of the Medtronic Overnight Closed Loop study is to assess the performance of a system designed to automatically infuse the correct insulin dose during the hours that the patient is sleeping.
The system consists of an insulin pump that provides insulin to the patient through an infusion set. A sensor inserted just under the patient's skin measures glucose levels and a transmitter sends this information to the pump. To enable the sensor to register the glucose information correctly, it must be set (calibrated) by a finger stick blood glucose 3-4 times a day. In the commercially released system, the physician would recommend the continuous background (basal) insulin rates and the patient would be required to make decisions regarding extra insulin (bolus) for meals or as a response to high glucose levels.
In the Closed Loop System, a mobile control system is added to these devices. This consists of an Android phone, a closed loop algorithm and a translator. This system is designed to translate the sensor information and direct the pump to provide the required dosage of insulin automatically without requiring input from the patient.
Reliable calibration has proved challenging, and so it is important that the system function safely, even when calibration is inaccurate. In this study, a calibration error will be introduced under very controlled circumstances. This testing will identify if the system can maintain acceptable overnight glucose levels, regardless of whether or not calibration is ideal.
研究设计
- 研究类型
- Interventional
- 分配方式
- Non Randomized
- 干预模型
- Crossover
- 主要目的
- Treatment
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 70 Years(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •A clinical diagnosis of Type 1 diabetes as determined by the Investigator for a minimum of 36 months prior to enrollment
- •Weight is between 50 and 120 kg
- •Subject has used a Medtronic insulin pump for at least the previous 6 months
- •Subject has an glycated hemoglobin (A1C) value ≤ 9.0% demonstrated at the time of enrollment
- •Subject uses a rapid-acting analogue insulin in his/her pump
排除标准
- •Female subject who has a positive serum pregnancy screening test, or who plans to become pregnant during the course of the study
- •Subject has a history of hypoglycemic seizure or hypoglycemic coma within the past 12 months
- •Subject has an exclusionary cardiac disorder
结局指标
主要结局
Amount of time spent in the target range defined as 3.8-8.8 mmol/l (70 to 160 mg/dL) YSI glucose during in-clinic overnight visits
时间窗: Two in-patient nights
Subjects will arrive at the research facility between 17:00 and 18:00. Plasma glucose will be measured and they will have their evening meal of no more than 80 g of carbohydrate. They will bolus for this meal using their usual Bolus Wizard (Medtronic) settings. After the meal, an IV catheter will be inserted for the collection of venous blood samples. The Mobile Control System will be connected and closed-loop control will be started approximately 4 hours after dinner. The subject will remain under closed-loop control overnight (approximately 11 p.m.-6 a.m.). During one of the Closed Loop nights the system will be challenged by a calibration error of 30%. At approximately 7 am (before breakfast), the Mobile Control System will be turned off and the patient will be provided with a breakfast (approximately 40-70 gm of carbohydrate) and will take a bolus according to the Bolus Wizard. The patient will return to their usual pump treatment with the MiniMed Paradigm Veo System.
次要结局
- Glucose Area under the curve (AUC) below 3.8 mmol/l (70 mg/dL ) YSI glucose obtained during in-clinic overnight visits(Two overnight closed loop sessions)
- Glucose AUC above 8.8mmol/l (160 mg/dL) YSI glucose obtained during in-clinic overnight visits(Two overnight closed loop sessions)