Non-Invasive Hemorrhage Monitoring: A Pilot Study to Assess the Feasibility and Tolerability of the AccuFlow Perfusion Sensor for the Detection of Intrapartum Hemorrhage

Megan Lord2 个研究点 分布在 1 个国家目标入组 25 人开始时间: 2020年5月20日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
25
试验地点
2
主要终点
Improving sensor design and analysis

研究概览

简要总结

Obstetric hemorrhage is one of the leading causes of maternal death worldwide. One of the challenges in management of hemorrhage is that young, healthy women compensate for blood loss via peripheral vasoconstriction, so they maintain their blood pressure and heart rate at normal levels even after experiencing significant blood loss. By the time vital sign abnormalities appear, interventions must be performed extremely rapidly to avoid organ damage and maternal death. Clinical methods of estimating blood loss in real time, such as visual estimation, are notoriously unreliable, and changes in laboratory testing such as hemoglobin levels lag hours behind actual blood loss. A tool which can detect and quantify blood loss in real time, before vital sign changes occur, has the potential to allow for earlier mobilization of resources and intervention in these cases, thus saving lives. This device is meant to detect changes in skin blood flow which reflect vasoconstriction. The investigators believe that this device, therefore, has the potential to be able to detect and quantify blood loss in real-time. However, as this novel device has never been used for this purpose, before undertaking a large clinical trial, the investigators feel it is necessary to perform a pilot study to assess the feasibility and tolerability of this device. The investigators plan to test this by asking 50 patients undergoing planned cesarean section to wear the device during their surgery. The device will collect skin perfusion measurements during the surgery, which will not be available to the operating team. The patients will also be asked to complete a survey regarding their experience wearing the device. The investigators will use this information to ensure that the device is transmitting interpretable data, that patients feel the device is tolerable during surgery, and to ensure that the device can be used in the operating room without any unforeseen logistical challenges which would need to be addressed in planning a larger trial. The investigators will perform a preliminary comparison of sensor readings to laboratory findings, to assist in planning a larger trial.

详细描述

Hemorrhage remains the leading cause of direct pregnancy-related death in both the developing world and the developed world. Young, healthy patients compensate for hemorrhage via peripheral vasoconstriction, delaying the emergence of vital sign abnormalities until 15-30% of total blood volume has been lost (1-2L in a term pregnant patient). A significant drop in blood pressure and marked tachycardia (>120bpm) are not seen until 30-40% of blood volume has been lost. After 40% loss, however, patients may rapidly decompensate, leaving a narrow window between the emergence of vital sign abnormalities and rapid clinical decompensation. In facilities where obstetricians and anesthesiologists are not in-house at all times and blood product availability is limited, this narrow window may prove catastrophic.

Furthermore, management of obstetrical blood loss presents many unique challenges. First, during delivery, when hemorrhage is expected and occurs acutely and visibly, blood is often mixed with amniotic fluid, making it difficult to establish actual blood loss. Second, laboratory measurement of hemoglobin concentrations can be falsely normal during an acute episode of bleeding, as the hemoglobin measured does not reflect actual hemoglobin concentration until several hours after the hemorrhage event, when equilibration occurs. Third, unlike other surgical patients, postpartum women are expected to experience ongoing vaginal bleeding after their delivery. Bleeding which is persistently slightly heavier than average can lead to significant blood loss over a period of several days, without either the patient or the staff caring for her recognizing the hemorrhage until the patient becomes symptomatic or vital sign instability occurs. The postpartum uterus can serve as a reservoir for retained blood and clot, allowing a patient to continue bleeding into her uterus after delivery without any external evidence of bleeding. In a thin patient this is frequently recognized on "fundal checks" where the size of the uterus is evaluated, but in obese patients the uterine fundus may be difficult to appreciate, and a significant volume of blood can accumulate in the uterus before it is recognized.

Due to the high risk of mortality related to postpartum hemorrhage and the difficulties providers face quantifying blood loss accurately after delivery,, it is imperative to develop a tool that can detect and quantify hemorrhage before vital sign changes occur. This tool should be simple to use, allowing healthcare providers at all levels to identify hemorrhage early and to mobilize appropriate resources. This will facilitate earlier intervention to treat blood loss before clinical decompensation. As stated, vasoconstriction is the earliest physiologic responses to acute hypovolemia6 and, as such, would be an appropriate target for such a tool.

The AccuFlow Sensor is a device that is based on Combined Heat-Flux and Temperature Sensor (CHFT+) technology developed in the heat transfer laboratory at Virginia Tech and produced by ThermanSENSE Corp (Roanoke, VA). The device, which consists of a heat flux sensor, a thin-film thermocouple, and a heater element, makes direct, quantitative measurements of perfusion occurring at the surface of exposed tissue in real-time. The sensor is less than 1mm thick, covers one square inch of skin, and adheres to the patient's skin using a piece of medical-grade silicone adhesive tape. The heater applies a small amount of heat to the skin surface, raising its temperature to approximately 39 degrees C. The heat flux sensor measures the rate at which the heat dissipates into the tissue, and uses this to calculate perfusion using the Pennes Bio-Heat Transfer Equation, which was developed to describe patterns of heat distribution in the human forearm and which has been the standard tool to predict temperature distributions in living human tissue since the 1950s.10 The sensor measures both surface perfusion and deep perfusion (1-2cm below the skin surface).

The primary theoretical source of patient injury or discomfort from the device is thermal injury from the heater element. Under typical circumstances, the device heats up to approximately 39 degrees Celsius (102 Fahrenheit, the temperature of a typical hot tub), and to prevent thermal injury, the device automatically shuts off should it reach a temperature of 45 C (113F, just above the typical hot-shower temperature of 110F). The device has been tested for comfort and tolerability on 20 healthy volunteers, who reported no discomfort while wearing the device or afterwards. Prior studies have shown no evidence of thermal injury in tissues exposed to a temperature of 39C for prolonged periods, and that exposure to temperatures of 45C causes mild thermal injury only after 150 minutes of constant exposure.11 As this device only heats up while it is taking a measurement, and never higher than 45C, the risk of a thermal injury while wearing the device for the duration of a cesarean section as planned during this study is extremely low. However, should the sensor become uncomfortable to the patient, it can easily be removed. To date, the sensor has been used to evaluate viability of organs intended for transplantation, by measuring blood flow at the surface of the organ.7 Currently, the sensor is being used to evaluate differences in tissue perfusion between pediatric patients with sickle cell anemia and healthy controls in an ongoing research study at Virginia Tech. Of note, the device meets FDA criteria for a nonsignificant risk device, and has been determined to be a nonsignificant risk device by the Institutional Review Board at Virginia Tech/Carilion Clinic for use in adult and pediatric patients at Carilion Clinic facilities. As mentioned before, the patent-pending AccuFlow device is provided by ThermaSENSE Corp. (Roanoke, VA) and a number of sensors will be contributed for the purpose of this study. The device has not yet been studied for detection of obstetric hemorrhage.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
Female
接受健康志愿者

入选标准

  • Age 18 years or older
  • English speaking
  • Undergoing planned cesarean section at Women and Infants Hospital
  • BMI on admission less than or equal to 35 kg/m2
  • Preoperative hemoglobin greater than or equal to 11.5g/dL on routine labs

排除标准

  • History of allergy or other adverse reaction to adhesives
  • Fever at the time of recruitment

结局指标

主要结局

Improving sensor design and analysis

时间窗: 6 months

We will use the sensor readings obtained from this study to determine how best to obtain and analyze sensor readings in a larger trial. This will include comparing different ratios of readings (both different points in time during the case and different measurement sites on the patient) to determine the ideal number and locations of sensors, and ideal method of data analysis, for a larger trial.

次要结局

  • Accuracy(6 months)
  • Device Tolerability(6 months.)