System Accuracy Evaluation of 2 CE-marked Blood Glucose Monitoring System
- Conditions
- Diabetes
- Registration Number
- NCT02710617
- Lead Sponsor
- Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm
- Brief Summary
The objective of this study is the evaluation of system accuracy following ISO 15197:2013 (E), clause 6.3 in which accuracy requirements and applicable test procedures for blood glucose monitoring systems intended for self-monitoring of blood glucose by patients are stipulated.
However in this study, system accuracy evaluation will be performed for Hemocue Glucose 201+ and Hemocue Glucose 201 RT which are both glucose monitoring systems for professional use only on behalf of the IDT with one reagent system lot for each BGMS.
Main goal is the evaluation of system accuracy of both of the hemocue Systems. (english)
- Detailed Description
Not available
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 109
- Male or female, with type 1 diabetes, type 2 diabetes or subjects without diabetes
- Signed informed consent form
- Minimum age of 18 years
- Subjects are legally competent and capable to understand character, meaning and consequences of the study.
- If blood glucose values < 80 mg/dl or > 300 mg/dl shall be measured after short term alteration in insulin therapy:
- Male or female with type 1 diabetes and intensified insulin therapy or insulin pump therapy.
- Signature of subjects to document consent with these procedures on informed consent form.
- Pregnancy or lactation period
- Severe acute disease (at the study physician's discretion)
- Severe chronic disease with potential risk during the test procedures (at the study physician's discretion)
- Current constitution that compromises the subject's capability to participate in the study (at the study physician's discretion)
- Being unable to give informed consent
- < 18 years
- Legally incompetent
- Being committed to an institution (e.g. psychiatric clinic)
- Language barriers potentially compromising an adequate compliance with study procedures
- Dependent on investigator or sponsor
- If blood glucose values < 80 mg/dl shall be measured after short term alteration in insulin therapy (concentration category 2, see Fehler! Verweisquelle konnte nicht gefunden werden.), subjects with type 1 diabetes, suffering from:
- Coronary heart disease
- Condition after myocardial infarction
- Condition after cerebral events
- Peripheral arterial occlusive disease
- Hypoglycemia unawareness
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- Not specified
- Primary Outcome Measures
Name Time Method System accuracy criteria (see description) For each subject, the experimental phase has an expected duration of up to 6 hours System accuracy criteria
* Following ISO 15197:2013 (E), the BGMS shall meet both of the following criteria:
* Criterion A: 95 % of the measured glucose values shall fall within either ± 15 mg/dl (0.83 mmol/l) of the average comparison measurement result at glucose concentrations \< 100 mg/dl (5.55 mmol/l) or within ± 15 % at glucose concentrations ≥ 100 mg/dl (5.55 mmol/l).
* Each lot shall pass acceptability criterion A.
* Criterion B: 99 % of individual glucose measured values shall fall within zones A and B of the Consensus Error Grid (CEG) for type 1 diabetes.
- Secondary Outcome Measures
Name Time Method
Related Research Topics
Explore scientific publications, clinical data analysis, treatment approaches, and expert-compiled information related to the mechanisms and outcomes of this trial. Click any topic for comprehensive research insights.
Trial Locations
- Locations (1)
Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm
🇩🇪Ulm, Germany
Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm🇩🇪Ulm, Germany