Clinical Trials/NCT03812471
NCT03812471
Completed
Not Applicable

Optimization of Fetal Biometry With 3D Ultrasound and Image Recognition

Philips Electronics Nederland B.V. acting through Philips CTO organization1 site in 1 country265 target enrollmentOctober 23, 2018

Overview

Phase
Not Applicable
Intervention
Not specified
Conditions
Pregnancy Related
Sponsor
Philips Electronics Nederland B.V. acting through Philips CTO organization
Enrollment
265
Locations
1
Primary Endpoint
Bi-parietal diameter in mm.
Status
Completed
Last Updated
3 years ago

Overview

Brief Summary

In the proposed clinical investigation, the investigational device (BabySize3D) will provide automated plane extraction and automated biometry measurements from 3D US volumes. The results obtained from the investigational software

Registry
clinicaltrials.gov
Start Date
October 23, 2018
End Date
March 17, 2023
Last Updated
3 years ago
Study Type
Interventional
Study Design
Parallel
Sex
Female

Investigators

Sponsor
Philips Electronics Nederland B.V. acting through Philips CTO organization
Responsible Party
Sponsor

Eligibility Criteria

Inclusion Criteria

  • evolutive pregnancy with one fetus with a Gestational Age between 16 and 30 weeks
  • aged between 18 and 65 years.
  • social welfare benefit recipient
  • willing and able to provide informed consent

Exclusion Criteria

  • under 18 years old
  • multiple pregnancy
  • over 18 years old and mentioned in French law articles L.1126-6 et L-1126-8,
  • fetal malformation or suspected fetal malformation
  • technical conditions that do not allow for standard fetal biometry (abdominal wall, fetal position, BMI \> 25 kg/m2)
  • fetal biometry non indicated in the standard pregnancy follow-up
  • unwilling or unable to provide informed consent (e. g. physical, mental disability or linguistic factor that compromises patient information)

Outcomes

Primary Outcomes

Bi-parietal diameter in mm.

Time Frame: Day 1

The bi-parietal diameter in mm will be obtained with three methods: (1) fully automated 3D volume processing and caliper placement, (2) manual 3D volume processing and caliper placement, (3) manual selection of the 2D view plane and manual caliper placement on the screen of the ultrasound system (corresponding to standard of care measurements). The values obtained with the three methods and the differences between these values will be reported.

Head circumference in mm

Time Frame: Day 1

The Head circumference in mm will be obtained with three methods: (1) fully automated 3D volume processing and caliper placement, (2) manual 3D volume processing and caliper placement, (3) manual selection of the 2D view plane and manual caliper placement on the screen of the ultrasound system (corresponding to standard of care measurements). The values obtained with the three methods and the differences between these values will be reported.

Femur length in mm

Time Frame: Day 1

The Femur length in mm will be obtained with three methods: (1) fully automated 3D volume processing and caliper placement, (2) manual 3D volume processing and caliper placement, (3) manual selection of the 2D view plane and manual caliper placement on the screen of the ultrasound system (corresponding to standard of care measurements). The values obtained with the three methods and the differences between these values will be reported

Abdominal circumference in mm

Time Frame: Day 1

The Abdominal circumference in mm will be obtained with three methods: (1) fully automated 3D volume processing and caliper placement, (2) manual 3D volume processing and caliper placement, (3) manual selection of the 2D view plane and manual caliper placement on the screen of the ultrasound system (corresponding to standard of care measurements). The values obtained with the three methods and the differences between these values will be reported.

Secondary Outcomes

  • Difference of scores of the 2D standard abdominal plane and abdominal plane extracted from the 3D volume, based on six criteria listed in (Salomon et al. Ultrasound Obstet Gynecol 2006; 27: 34-40)(Day 1)
  • Intraclass correlation coefficient of the head circumference in mm (different operators)(Day 1)
  • Bland-Altman representation of the head circumference in mm (different operators)(Day 1)
  • Failure rate of the automated measurements(Day 1)
  • Exam and post-processing duration(Day 1)
  • Difference of scores of the 2D standard cephalic plane and cephalic plane extracted from the 3D volume, based on six criteria listed in (Salomon et al. Ultrasound Obstet Gynecol 2006; 27: 34-40)(Day 1)
  • Difference of scores of (i) the 2D standard abdominal plane and of (ii) the frames of a sequence of 2D images including the fetus abdomen (cine-loop), based on six criteria listed in (Salomon et al. Ultrasound Obstet Gynecol 2006; 27: 34-40)(Day 1)
  • Difference of scores of (i) the 2D standard femoral plane and of (ii) the frames of a sequence of 2D images including the fetus femur (cine-loop), based on four criteria listed in (Salomon et al. Ultrasound Obstet Gynecol 2006; 27: 34-40)(Day 1)
  • Intraclass correlation coefficient of the bi-parietal diameter in mm (different operators)(Day 1)
  • Bland-Altman representation of the head circumference in mm (same operator)(Day 1)
  • Difference of scores of the 2D standard femoral plane and femoral plane extracted from the 3D volume, based on four criteria listed in (Salomon et al. Ultrasound Obstet Gynecol 2006; 27: 34-40)(Day 1)
  • Difference of scores of (i) the 2D standard cephalic plane and of (ii) the frames of a sequence of 2D images including the fetus head (cine-loop), based on six criteria listed in (Salomon et al. Ultrasound Obstet Gynecol 2006; 27: 34-40)(Day 1)
  • Title: Intraclass correlation coefficient of the bi-parietal diameter in mm (same operator)(Day 1)
  • Bland-Altman representation of the bi-parietal diameter in mm (same operator)(Day 1)
  • Bland-Altman representation of the bi-parietal diameter in mm (different operators)(Day 1)
  • Intraclass correlation coefficient of the head circumference in mm (same operator)(Day 1)
  • Intraclass correlation coefficient of the abdominal circumference in mm (same operator)(Day 1)
  • Bland-Altman representation of the abdominal circumference in mm (same operator)(Day 1)
  • Intraclass correlation coefficient of the femoral length in mm (same operator)(Day 1)
  • Intraclass correlation coefficient of the abdominal circumference in mm (different operators)(Day 1)
  • Bland-Altman representation of the abdominal circumference in mm (different operators)(Day 1)
  • Bland-Altman representation of the femoral length in mm (same operator)(Day 1)
  • Title: Intraclass correlation coefficient of the femoral length in mm (different operators)(Day 1)
  • Bland-Altman representation of the femoral length in mm (different operators)(Day 1)